info:eu-repo/semantics/article
Fluorimetric determination of fluoride in a flow assembly integrated on-line to an open/closed FIA system to remove interference by solid phase extraction
Fecha
2002-11Registro en:
Garrido, Mariano Enrique; Lista, Adriana Guillermina; Palomeque, Miriam Edid; Fernández Band, Beatriz Susana; Fluorimetric determination of fluoride in a flow assembly integrated on-line to an open/closed FIA system to remove interference by solid phase extraction; Elsevier Science; Talanta; 58; 5; 11-2002; 849-853
0039-9140
CONICET Digital
CONICET
Autor
Garrido, Mariano Enrique
Lista, Adriana Guillermina
Palomeque, Miriam Edid
Fernández Band, Beatriz Susana
Resumen
A simple flow injection fluorimetric method for fluoride determination is proposed. The method is based on the enhanced fluorescence of quercitin-Zr(IV) complex when fluoride ion is present in the sample. An open/closed FIA manifold with a mini-column of Dowex 50W X8 resin was used to remove the most important interference (aluminum). The two FIA assemblies were integrated on-line to automate the pretreatment of the water sample and fluoride determination. The calibration graph was linear over the range 0.1-3.0 μg ml-1 of fluoride with a correlation coefficient of 0.999 and LOD 0.06 μg ml-1. The relative standard deviation was 2.5% and the sample throughput was 52 h-1 without pretreatment and 10 h-1 with pretreatment of the sample. The method was applied to the determination of fluoride in water samples.