dc.creatorGastao Cruz
dc.creatorAlina Hua
dc.creatorCamila Munoz
dc.creatorTevfik Ismail
dc.creatorAmedeo Chiribiri
dc.creatorRene Michael Botnar
dc.creatorClaudia Prieto
dc.date.accessioned2024-03-04T15:19:42Z
dc.date.available2024-03-04T15:19:42Z
dc.date.created2024-03-04T15:19:42Z
dc.date.issued2023
dc.identifier10.1002/mrm.29626
dc.identifier1522-2594
dc.identifier0740-3194
dc.identifierhttps://doi.org/10.1002/mrm.29626
dc.identifierhttps://repositorio.uc.cl/handle/11534/82470
dc.description.abstractPurposeDevelop a novel approach for accelerated 2D free-breathing myocardial perfusion via low-rank motion-corrected (LRMC) reconstructions.MethodsMyocardial perfusion imaging requires high spatial and temporal resolution, despite scan time constraints. Here, we incorporate LRMC models into the reconstruction-encoding operator, together with high-dimensionality patch-based regularization, to produce high quality, motion-corrected myocardial perfusion series from free-breathing acquisitions. The proposed framework estimates beat-to-beat nonrigid respiratory (and any other incidental) motion and the dynamic contrast subspace from the actual acquired data, which are then incorporated into the proposed LRMC reconstruction. LRMC was compared with iterative SENSitivity Encoding (SENSE) (itSENSE) and low-rank plus sparse (LpS) reconstruction in 10 patients based on image-quality scoring and ranking by two clinical expert readers.ResultsLRMC achieved significantly improved results relative to itSENSE and LpS in terms of image sharpness, temporal coefficient of variation, and expert reader evaluation. Left ventricle image sharpness was approximately 75%, 79%, and 86% for itSENSE, LpS and LRMC, respectively, indicating improved image sharpness for the proposed approach. Corresponding temporal coefficient of variation results were 23%, 11% and 7%, demonstrating improved temporal fidelity of the perfusion signal with the proposed LRMC. Corresponding clinical expert reader scores (1–5, from poor to excellent image quality) were 3.3, 3.9 and 4.9, demonstrating improved image quality with the proposed LRMC, in agreement with the automated metrics.ConclusionLRMC produces motion-corrected myocardial perfusion in free-breathing acquisitions with substantially improved image quality when compared with iterative SENSE and LpS reconstructions.
dc.languageen
dc.rightsCC BY 4.0 DEED Attribution 4.0 International
dc.rightshttps://creativecommons.org/licenses/by/4.0/
dc.rightsacceso abierto
dc.subjectAccelerated scan
dc.subjectCardiac MR
dc.subjectFree breathing
dc.subjectLow rank
dc.subjectMotion correction
dc.subjectMyocardial perfusion
dc.titleLow-rank motion correction for accelerated free-breathing first-pass myocardial perfusion imaging
dc.typeartículo