Evaluation of equivalent and effective dose by KAP for patient and orthopedic surgeon in vertebral compression fracture surgery
Registro en:
0969-806X
132
10.1016/j.radphyschem.2016.11.011
43.267
Autor
SANTOS, FELIPE A.
GALEANO, DIEGO C.
SANTOS, WILLIAM S.
SILVA, ADEMIR X.
SOUZA, SUSANA O.
CARVALHO JUNIOR, ALBERICO B.
Resumen
Clinical scenarios were virtually modeled to estimate both the equivalent and effective doses normalized by KAP (Kerma Area Product) to vertebra compression fracture surgery in patient and surgeon. This surgery is known as kyphoplasty and involves the use of X-ray equipment, the C-arm, which provides real-time images to assist the surgeon in conducting instruments inserted into the patient and in the delivery of surgical cement into the fractured vertebra. The radiation transport code used was MCNPX (Monte Carlo N-Particle eXtended) and a pair of UFHADM (University of Florida Hybrid ADult Male) virtual phantoms. The developed scenarios allowed us to calculate a set of equivalent dose (H-T) and effective dose (E) for patients and surgeons. In additional, the same scenario was calculated KAP in the tube output and was used for calculating conversion coefficients (E/KAP and H-T/KAP). From the knowledge of the experimental values of KAP and the results presented in this study, it is possible to estimate absolute values of effective doses for different exposure conditions. In this work, we developed scenarios with and without the surgical table with the purpose of comparison with the existing data in the literature. The absence of the bed in the scenario promoted a percentage absolute difference of 56% in the patient effective doses in relation to scenarios calculated with a bed. Regarding the surgeon, the use of the personal protective equipment (PPE) reduces between 75% and 79% the effective dose and the use of the under table shield (UTS) reduces the effective dose of between 3% and 7%. All these variations emphasize the importance of the elaboration of virtual scenarios that approach the actual clinical conditions generating E/KAP and H-T/KAP closer to the actual values.