dc.contributorCalvo-Vargas, C., University of Guadalajara, Department of Internal Medicine, Hospital Civil Juan I. Menchaca, Guadalajara, Mexico, University of Guadalajara, Physiology Department, Health Sciences Center, Guadalajara, Mexico, Pablo B. Villase�or 387 SH, Guadalajara, Jal., Mexico; Padilla-Rios, V., University of Guadalajara, Department of Internal Medicine, Hospital Civil Juan I. Menchaca, Guadalajara, Mexico; Troyo-Sanrom�n, R., University of Guadalajara, Physiology Department, Health Sciences Center, Guadalajara, Mexico
dc.creatorCalvo-Vargas, C.
dc.creatorPadilla-Rios, V.
dc.creatorTroyo-Sanroman, R.
dc.date.accessioned2015-09-15T18:19:07Z
dc.date.accessioned2023-07-03T22:56:02Z
dc.date.available2015-09-15T18:19:07Z
dc.date.available2023-07-03T22:56:02Z
dc.date.created2015-09-15T18:19:07Z
dc.date.issued2003
dc.identifierhttp://www.scopus.com/inward/record.url?eid=2-s2.0-0037636365&partnerID=40&md5=af057308b5e1f8a7f95b53426e6c3ead
dc.identifierhttp://hdl.handle.net/20.500.12104/42579
dc.identifier10.1097/00126097-200304000-00002
dc.identifier.urihttps://repositorioslatinoamericanos.uchile.cl/handle/2250/7247480
dc.description.abstractObjective: To compare the correspondence between the loaned self-measurement equipment model (LSEM), 24 h awake ambulatory blood pressure monitoring (ABPM) and clinic blood pressure (CBP). Design: A cross-sectional study was carried out involving 90 hypertensive and normotensive subjects. Methods: We used a variation of home measurement, the loaned self-measurement equipment model (LSEM), that consisted of lending a number of sphygmomanometers, the property of the clinic, to patients for 3-day periods. The correspondence of results from LSEM was measured against the difference between the two averages, the correlation coefficients and the average unweighted absolute differences. Results: The average difference between CBP and ABPM was 6.1/2.5 mmHg (P<0.001, r = 0.54/0.63). Between CBP and LSEM, it was 6.6/1.4 mmHg (P<0.001/ 0.08, r = 0.52/0.63), and between the ABPM and LSEM 0.5/-1.0 mmHg (P = 0.70/0.14, r = 0.71/0.75). The average unweighted absolute differences between LSEM and ABPM readings were 5.5 and 1.5 mmHg for systolic and diastolic blood pressure, respectively, whereas the average difference between CBP and ABPM means was much higher, at 6.8 mmHg for systolic and 2.7 mmHg for diastolic pressure. Agreement in the diagnosis of systolic hypertension between LSEM and ABPM was found in 78.8% of the individuals, and of diastolic hypertension in 76.7%. Conclusion: Blood pressure readings taken using LSEM have a better correspondence than CBP in predicting the awake ABPM value, although the disagreement in terms of diagnosing hypertension should be noted. � 2003 Lippincott Williams & Wilkins.
dc.relationScopus
dc.relationBlood Pressure Monitoring
dc.relation8
dc.relation2
dc.relation63
dc.relation70
dc.titleLoaned self-measurement equipment model compared with ambulatory blood pressure monitoring