Artículos de revistas
A comparative study of microstrain around three-morse taper implants with machined and plastic copings under axial loading
Fecha
2013-01-01Registro en:
Brazilian Journal of Oral Sciences, v. 9, n. 1, p. 11-15, 2013.
1677-3217
1677-3225
2-s2.0-84891485064
Autor
Universidade Estadual de Campinas (UNICAMP)
Universidade Estadual Paulista (UNESP)
Institución
Resumen
Aim: To evaluate the occurrence of microstrain around morse taper implants in straight configuration under axial load in a cast monoblock framework. Methods: Three implants were inserted in a polyurethane block and microunit abutments were installed on the implants with 20 Ncm torque. Plastic and machined copings were adapted on the preset waxing to fabricate the framework (n=5). Four strain gauges were attached on the upper surface of the block and then each framework was tightened on the abutments and a vertical load of 30 kg was applied to five points of the framework. Results: The data obtained in the strain gauge analysis were subjected to twoway ANOVA and Tukey's test (α=0.05). There was statistically significant difference (p=0.0222) for the factor application point and the mean microstrain values were: application point B 402,04μσ, point A 401.21μσ, point E 390.44μσ , point D 341.76μσ and point C 309.19 μσ. Conclusions: There was no microstrain difference between plastic and machined copings during axial loading. Difference in the application point was observed, but remained within bone physiological limits.