Digital Image Correlation and Finite Element Analysis of Bone Strain Generated by Implant-Retained Cantilever Fixed Prosthesis
Fecha
2020-02-27Registro en:
The European journal of prosthodontics and restorative dentistry, v. 28, n. 1, p. 10-17, 2020.
0965-7452
10.1922/EJPRD_1941Tribst08
2-s2.0-85081075432
0056660842284397
0000-0002-1458-601X
Autor
Universidade Estadual Paulista (Unesp)
Clinic for Fixed and Removable Prosthodontics and Dental Materials Science
Institución
Resumen
PURPOSE: The present study evaluated the displacement and strain generated in an implant- supported fixed prosthesis under axial and non-axial loads using two methods. MATERIALS AND METHODS: Three implants were inserted in a resin block. The Digital Image Correlation (DIC) was used to measure displacement and strain generated on the surface of the resin blocks for the different load applications (500N, 1 image/second). A 3-dimensional model was constructed and a load of 500 N was applied at an axial point and a non-axial point through finite element analysis (FEA). RESULTS: Both methods gave similar trends for the strains, and both gave slightly higher strains with non-axial loading. FEA predicted higher strain magnitude (±11%) in comparison with DIC, but with the same mechanical behavior. According to ANOVA, the loading influenced the strain concentration. Higher strain was generated for non-axial loading around the implant nearest to the loading. CONCLUSIONS: For implant-retained cantilever fixed prosthesis, the same load applied in the lever arm induces higher strain in the cervical area of the last implant, which suggests more damaging potential than a load applied at the center of the prosthesis.