info:eu-repo/semantics/article
Water-equivalence of gel dosimeters for radiology medical imaging
Fecha
2018-11Registro en:
Valente, Mauro Andres; Vedelago, José Alberto; Chacón Obando, David; Mattea, Facundo; Velásquez, José; et al.; Water-equivalence of gel dosimeters for radiology medical imaging; Pergamon-Elsevier Science Ltd; Applied Radiation and Isotopes; 141; 11-2018; 193-198
0969-8043
CONICET Digital
CONICET
Autor
Valente, Mauro Andres
Vedelago, José Alberto
Chacón Obando, David
Mattea, Facundo
Velásquez, José
Pérez, Pedro Antonio
Resumen
International dosimetry protocols are based on determinations of absorbed dose to water. Ideally, the phantom material should be water equivalent; that is, it should have the same absorption and scatter properties as water. This study presents theoretical, experimental and Monte Carlo modeling of water-equivalence of Fricke and polymer (NIPAM, PAGAT and itaconic acid ITABIS) gel dosimeters. Mass and electronic densities along with effective atomic number were calculated by means of theoretical approaches. Samples were scanned by standard computed tomography. Photon mass attenuation coefficients and electron stopping powers were examined. Theoretical, Monte Carlo and experimental results confirmed good water-equivalence for all gel dosimeters. Overall variations with respect to water in the low energy radiology range (up to 130 kVp) were found to be less than 3% in average.