info:eu-repo/semantics/article
New, inexpensive and simple 3D printable device for nephelometric and fluorimetric determination based on smartphone sensing
Fecha
2020-05Registro en:
Vidal, Ezequiel Martin; Lorenzetti, Anabela Silvana; Aguirre, Miguel Ángel; Canals, Antonio; Domini, Claudia Elizabeth; New, inexpensive and simple 3D printable device for nephelometric and fluorimetric determination based on smartphone sensing; Royal Society of Chemistry; RSC Advances; 10; 33; 5-2020; 19713-19719
2046-2069
CONICET Digital
CONICET
Autor
Vidal, Ezequiel Martin
Lorenzetti, Anabela Silvana
Aguirre, Miguel Ángel
Canals, Antonio
Domini, Claudia Elizabeth
Resumen
A new, inexpensive and easy to use 3D printable device was developed for nephelometric and fluorimetric determination. Its applicability was tested for the quantification of quinine in tonic drinks and sulfate in natural water with good analytical accuracy. In this way, sulfate determination was carried out by nephelometry using a red LED, while quinine was determined using a blue LED by fluorimetry. A smartphone camera was used to take the pictures and afterwards transform them into the RGB color space using the software ImageJ by a personal computer. The linear range was 2.0-50.0 mg L-1for sulfate with a LOD of 0.13 mg L-1, and the corresponding quantification limit (LOQ) was 0.43 mg L-1. The linear range for quinine was from 0.42 to 3.10 mg L-1. The LOD and LOQ were 0.11 mg L-1and 0.38 mg L-1, respectively.