dc.description.abstractA passive neutron area monitor has been designed using Monte Carlo methods; the monitor is a polyethylene cylinder with pairs of thermoluminescent dosimeters (TLD600 and TLD700) as thermal neutron detector. The monitor was calibrated with a bare and a thermalzed 241AmBe neutron sources and its performance was evaluated measuring the ambient dose equivalent due to photoneutrons produced by a 15 MV linear accelerator for radiotherapy and the neutrons in the output of a TRIGA Mark III radial beam port.
dc.languageeng
dc.publisherElsevier
dc.relationgeneralPublic
dc.relationhttps://www.sciencedirect.com/science/article/abs/pii/S1350448714002406
dc.rightshttp://creativecommons.org/licenses/by-nc-sa/3.0/us/
dc.rightsAtribución-NoComercial-CompartirIgual 3.0 Estados Unidos de América
dc.sourceRadiation Measurements Vol. 69, octubre 2014, pp. 30-34
dc.titlePassive neutron area monitor with pairs of TLDs as neutron detector