dc.descriptionA SEMI-AUTOMATED INSTRUMENT FOR MEASURING HEAT TRANSFER PARAMETERS AROUND A HORIZONTAL CYLINDER IS PRESENTED. THE METHOD UPON WHICH THE INSTRUMENT IS BASED TAKES ADVANTAGE OF THE INTERFERENCE PHENOMENON AND A DIGITAL PC. THE INTERFEROGRAMS OBTAINED ARE DIGITALLY PRE-PROCESSED IN ORDER TO ENHANCE THE FRINGES AND CYLINDER SURFACE DEFINITIONS. AFTER APPLYING ENHANCING ALGORITHMS, A COMPUTING ALGORITHM IS APPLIED EVERY 45 DEGREES AROUND THE CYLINDER CONTOUR FOR GETTING THE LOCAL NUSSELT NUMBER AND TEMPERATURE VALUES. COMPARING WITH OTHER AUTHOR'S WORKS [17,19,21J, WE HAVE ERRORS LESS THAN 10% IN THE COMPUTATION OF BOTH THE NUSSELT NUMBER AND THE TEMPERATURE VALUES THE PRESENTED WORK IS INTENDEDTO BE THE FIRST STAGE FOR AFULL-AUTOMATED HEAT TRANSFER MEASURING INSTRUMENT.en-US
dc.descriptionA SEMI-AUTOMATED INSTRUMENT FOR MEASURING HEAT TRANSFER PARAMETERS AROUND A HORIZONTAL CYLINDER IS PRESENTED. THE METHOD UPON WHICH THE INSTRUMENT IS BASED TAKES ADVANTAGE OF THE INTERFERENCE PHENOMENON AND A DIGITAL PC. THE INTERFEROGRAMS OBTAINED ARE DIGITALLY PRE-PROCESSED IN ORDER TO ENHANCE THE FRINGES AND CYLINDER SURFACE DEFINITIONS. AFTER APPLYING ENHANCING ALGORITHMS, A COMPUTING ALGORITHM IS APPLIED EVERY 45 DEGREES AROUND THE CYLINDER CONTOUR FOR GETTING THE LOCAL NUSSELT NUMBER AND TEMPERATURE VALUES. COMPARING WITH OTHER AUTHOR'S WORKS [17,19,21J, WE HAVE ERRORS LESS THAN 10% IN THE COMPUTATION OF BOTH THE NUSSELT NUMBER AND THE TEMPERATURE VALUES THE PRESENTED WORK IS INTENDEDTO BE THE FIRST STAGE FOR AFULL-AUTOMATED HEAT TRANSFER MEASURING INSTRUMENT.es-ES
dc.formatapplication/pdf
dc.languagespa
dc.publisherComputación y Sistemases-ES
dc.relationhttp://ojs.unam.mx/index.php/cys/article/view/2526/2088
dc.sourceComputación y Sistemas; Vol 3, No 003 (2000)es-ES
dc.source1405-5546
dc.subjectHEAT TRANSFER MEASUREMENT; NUSSELT NUMBER; INTERFEROGRAM PROCESSINGen-US