dc.creatorSILVA, J.O.
dc.creatorCALDAS, L.V.E.
dc.creatorINTERNATIONAL JOINT CONFERENCE RADIO 2017
dc.date2018-02-26T17:04:41Z
dc.date2018-02-26T17:04:41Z
dc.dateSeptember 24-29, 2017
dc.date.accessioned2023-09-28T14:06:24Z
dc.date.available2023-09-28T14:06:24Z
dc.identifierhttp://repositorio.ipen.br/handle/123456789/28579
dc.identifier.urihttps://repositorioslatinoamericanos.uchile.cl/handle/2250/8998829
dc.descriptionA special mammography homemade ionization chamber was developed to be applied for mammography energy range dosimetry. This chamber has a total sensitive volume of 6 cm3 and is made of a PMMA body and graphite coated collecting electrode. Performance tests as saturation, ion collection efficiency, linearity of chamber response versus air kerma rate and energy dependence were determined. The results obtained with this special homemade ionization chamber are within the limits stated in international recommendations. This chamber can be used in quality control programs of mammography energy range. All measurements were carried out at the Calibration Laboratory of IPEN.
dc.format1-3
dc.publisherSociedade Brasileira de Prote????o Radiol??gica
dc.rightsopenAccess
dc.subjectdosimetry
dc.subjectionization chambers
dc.subjectmammary glands
dc.subjectperformance testing
dc.subjectpmma
dc.subjectquality control
dc.titlePerformance tests of a special ionization chamber for X-rays in mammography energy range
dc.typeTexto completo de evento
dc.coverageI
dc.localMaring??, PR


Este ítem pertenece a la siguiente institución