dc.creatorCASTRO, MAYSA C. de
dc.creatorXAVIER, MARCOS
dc.creatorCALDAS, LINDA V.E.
dc.creatorBRAZILIAN CONGRESS ON METROLOGY, 8th
dc.date2017-03-20T11:45:32Z
dc.date2017-03-20T11:45:32Z
dc.date29 de novembro - 04 de dezembro, 2015
dc.date.accessioned2023-09-28T13:32:57Z
dc.date.available2023-09-28T13:32:57Z
dc.identifierhttp://repositorio.ipen.br/handle/123456789/27218
dc.identifier.urihttps://repositorioslatinoamericanos.uchile.cl/handle/2250/8997496
dc.descriptionThe use of computed tomography (CT) has increased over the years, thus generating a concern about the doses received by patients undergoing this procedure. Therefore, it is necessary to perform routinely beam dosimetry with the use of a pencil-type ionization chamber. This detector is the most utilized in the procedures of quality control tests on this kind of equipment. The objective of this work was to perform some characterization tests in standard CT beams, as the saturation curve, polarity effect, ion collection efficiency and linearity of response, using three ionization chambers, one commercial and two developed at the IPEN.
dc.format1-5
dc.publisherBristol, UK: IOP Publishing
dc.rightsopenAccess
dc.sourceJournal of Physics: Conference Series, v. 733, 2016
dc.subjectbeams
dc.subjectcomparative evaluations
dc.subjectcomputerized tomography
dc.subjectdosimetry
dc.subjectionization chambers
dc.subjectbrazilian cnen
dc.subjectperformance
dc.subjectquality control
dc.subjectsaturation
dc.titlePerformance of three pencil-type ionization chambers (10 cm) in computed tomography standard beams
dc.typeTexto completo de evento
dc.coverageI