Otro
Comparison of Single-Standing or Connected Implants on Stress Distribution in Bone of Mandibular Overdentures: A Two-Dimensional Finite Element Analysis
Registro en:
Journal of Craniofacial Surgery. Philadelphia: Lippincott Williams & Wilkins, v. 21, n. 3, p. 696-702, 2010.
1049-2275
10.1097/SCS.0b013e3181d7f202
WOS:000278102200016
Autor
Tabata, Lucas Fernando
Assunção, Wirley Goncalves
Ricardo Barao, Valentim Adelino
Gomes, Erica Alves
Delben, Juliana Aparecida
Sousa, Edson Antonio Capello
Rocha, Eduardo Passos
Resumen
This study aimed to compare the influence of single-standing or connected implants on stress distribution in bone of mandibular overdentures by means of two-dimensional finite element analysis. Two finite element models were designed using software (ANSYS) for 2 situations: bar-clip (BC) group-model of an edentulous mandible supporting an overdenture over 2 connected implants with BC system, and o'ring (OR) group-model of an edentulous mandible supporting an overdenture over 2 single-standing implants with OR abutments. Axial loads (100 N) were applied on either central (L1) or lateral (L2) regions of the models. Stress distribution was concentrated mostly in the cortical bone surrounding the implants. When comparing the groups, BC (L1, 52.0 MPa and L2, 74.2 MPa) showed lower first principal stress values on supporting tissue than OR (L1, 78.4 MPa and L2, 76.7 MPa). Connected implants with BC attachment were more favorable on stress distribution over peri-implant-supporting tissue for both loading conditions.