dc.contributorJakubiak, Rosangela Requi
dc.contributorhttp://lattes.cnpq.br/9416396245605040
dc.contributorJakubiak, Rosangela Requi
dc.contributorMiquelin, Charlie Antoni
dc.contributorCarvalho Neto, Arnolfo de
dc.creatorOlimpio, Aline Beatriz
dc.date.accessioned2019-07-23T18:17:52Z
dc.date.accessioned2022-12-06T15:12:00Z
dc.date.available2019-07-23T18:17:52Z
dc.date.available2022-12-06T15:12:00Z
dc.date.created2019-07-23T18:17:52Z
dc.date.issued2019-03-29
dc.identifierOLIMPIO, Aline Beatriz. Controle de qualidade em espectroscopia de prótons por ressonância magnética. 2019. 86 f. Dissertação (Mestrado em Engenharia Biomédica) - Universidade Tecnológica Federal do Paraná, Curitiba, 2019.
dc.identifierhttp://repositorio.utfpr.edu.br/jspui/handle/1/4219
dc.identifier.urihttps://repositorioslatinoamericanos.uchile.cl/handle/2250/5262246
dc.description.abstractThis study presents the development of a quality control routine for proton magnetic resonance spectroscopy performed on the Siemens Avanto Fit 1,5 T equipment. The objective of this study is to evaluate the repeatability of the system in the acquisition of single and multi voxel spectroscopy, using the STEAM (Stimulated Echo Acquisition Mode) and SE (spin echo) techniques, also called PRESS (Point Resolved Spectroscopy), analyzing data from two metabolites: Acetate and Lactate, in relation to peak amplitude and integral below the peak area, for short and long echo times. The phantom D170 Spectroscopy was positioned on the 18-channel head matrix radio frequency coil, the manufacturer's quality control protocol was used. A total of 15 samples were obtained on two different dates, being 8 samples on the first day, a 30 day interval and 7 samples on the second day, after the preventive maintenance of the equipment, ensuring that the other parameters are within the normal range. The information was collected, the data tabulated and analyzed graphically. The test of homogeneity of the magnetic field had results below the values established by the regulatory standards, which guarantees the accuracy in the identification of the peak of frequency of the water, for later application of the technique of water suppression. It was observed that the SE / PRESS technique was more advantageous because it has a higher signal-to-noise ratio, but the STEAM technique may be preferable in noncollaborative patients. From the established methodology, since the manufacturer does not provide reference values, the repeatability test was applied so that the spectral quality control was acquired, reference intervals were established. With these results, the routine can be inserted in the routine of periodic tests, the values of baseline can be defined for the other equipment of the clinic. This determination of the equipment reference results in greater reliability in the method and, consequently, in the diagnosis resulting from it.
dc.publisherUniversidade Tecnológica Federal do Paraná
dc.publisherCuritiba
dc.publisherBrasil
dc.publisherPrograma de Pós-Graduação em Engenharia Biomédica
dc.publisherUTFPR
dc.rightsopenAccess
dc.subjectEspectroscopia
dc.subjectDiagnóstico por imagem
dc.subjectControle de qualidade
dc.subjectAnálise espectral
dc.subjectCampos magnéticos - Testes
dc.subjectImagem de ressonância magnética - Avaliação
dc.subjectConfiabilidade
dc.subjectInstrumentos e aparelhos médicos - Calibração
dc.subjectFísica médica
dc.subjectEngenharia biomédica
dc.subjectSpectroscopy
dc.subjectDiagnostic imaging
dc.subjectQuality control
dc.subjectSpectrum analysis
dc.subjectMagnetic fields - Testing
dc.subjectMagnetic resonance imaging - Evaluation
dc.subjectReliability
dc.subjectMedical instruments and apparatus - Calibration
dc.subjectMedical physics
dc.subjectBiomedical engineering
dc.titleControle de qualidade em espectroscopia de prótons por ressonância magnética
dc.typemasterThesis


Este ítem pertenece a la siguiente institución