dc.creatorDiaz, Susana Beatriz
dc.creatorBooth, Charles R.
dc.creatorArmstrong, Roy
dc.creatorBrunat, Claudio
dc.creatorCabrera, Sergio
dc.creatorCamilion, Carolina
dc.creatorCasiccia, Claudio
dc.creatorDeferrari, Guillermo Alejandro
dc.creatorFuenzalida, Humberto
dc.creatorLovengreen, Charlotte
dc.creatorPaladini, Alejandro Alberto
dc.creatorPedroni, Jorge
dc.creatorRosales, Alejandro
dc.creatorZagarese, Horacio Ernesto
dc.creatorVernet, Maria
dc.date.accessioned2019-08-05T23:06:19Z
dc.date.accessioned2022-10-15T15:33:44Z
dc.date.available2019-08-05T23:06:19Z
dc.date.available2022-10-15T15:33:44Z
dc.date.created2019-08-05T23:06:19Z
dc.date.issued2005-09
dc.identifierDiaz, Susana Beatriz; Booth, Charles R.; Armstrong, Roy; Brunat, Claudio; Cabrera, Sergio; et al.; Multichannel radiometer calibration: A new approach; Optical Society of America; Applied Optics; 44; 26; 9-2005; 5374-5380
dc.identifier1539-4522
dc.identifierhttp://hdl.handle.net/11336/80959
dc.identifier2155-3165
dc.identifierCONICET Digital
dc.identifierCONICET
dc.identifier.urihttps://repositorioslatinoamericanos.uchile.cl/handle/2250/4403470
dc.description.abstractThe error in irradiance measured with Sun-calibrated multichannel radiometers may be large when the solar zenith angle (SZA) increases. This could be particularly detrimental in radiometers installed at mid and high latitudes, where SZAs at noon are larger than 50° during part of the year. When a multi-regressive methodology, including the total ozone column and SZA, was applied in the calculation of the calibration constant, an important improvement was observed. By combining two different equations, an improvement was obtained at almost all the SZAs in the calibration. An independent test that compared the irradiance of a multichannel instrument and a spectroradiometer installed in Ushuaia, Argentina, was used to confirm the results.
dc.languageeng
dc.publisherOptical Society of America
dc.relationinfo:eu-repo/semantics/altIdentifier/doi/http://dx.doi.org/10.1364/AO.44.005374
dc.relationinfo:eu-repo/semantics/altIdentifier/url/https://www.osapublishing.org/ao/abstract.cfm?uri=ao-44-26-5374
dc.rightshttps://creativecommons.org/licenses/by-nc-sa/2.5/ar/
dc.rightsinfo:eu-repo/semantics/restrictedAccess
dc.subjectmultichannel radiometer calibration
dc.subjectirradiance measured
dc.titleMultichannel radiometer calibration: A new approach
dc.typeinfo:eu-repo/semantics/article
dc.typeinfo:ar-repo/semantics/artículo
dc.typeinfo:eu-repo/semantics/publishedVersion


Este ítem pertenece a la siguiente institución