dc.identifierParedes Mellone, Oscar Ariel; Bianco, Leonardo Martín; Ceppi, Sergio Andres; Goncalves Honnicke, M.; Stutz, Guillermo Eduardo; Background radiation in inelastic X-ray scattering and X-ray emission spectroscopy. A study for Johann-type spectrometers; Elsevier Science; Nuclear Instruments and Methods in Physics Research A: Accelerators, Spectrometers, Detectors and Associated Equipament; 894; 21-6-2018; 119-128
dc.identifier0168-9002
dc.identifierhttp://hdl.handle.net/11336/91564
dc.identifierCONICET Digital
dc.identifierCONICET
dc.identifier.urihttps://repositorioslatinoamericanos.uchile.cl/handle/2250/4357136
dc.description.abstractA study of the background radiation in inelastic X-ray scattering (IXS) and X-ray emission spectroscopy (XES) based on an analytical model is presented. The calculation model considers spurious radiation originated from elastic and inelastic scattering processes along the beam paths of a Johann-type spectrometer. The dependence of the background radiation intensity on the medium of the beam paths (air and helium), analysed energy and radius of the Rowland circle was studied. The present study shows that both for IXS and XES experiments the background radiation is dominated by spurious radiation owing to scattering processes along the sample-analyser beam path. For IXS experiments the spectral distribution of the main component of the background radiation shows a weak linear dependence on the energy for the most cases. In the case of XES, a strong non-linear behaviour of the background radiation intensity was predicted for energy analysis very close to the backdiffraction condition, with a rapid increase in intensity as the analyser Bragg angle approaches π∕2. The contribution of the analyser–detector beam path is significantly weaker and resembles the spectral distribution of the measured spectra. Present results show that for usual experimental conditions no appreciable structures are introduced by the background radiation into the measured spectra, both in IXS and XES experiments. The usefulness of properly calculating the background profile is demonstrated in a background subtraction procedure for a real experimental situation. The calculation model was able to simulate with high accuracy the energy dependence of the background radiation intensity measured in a particular XES experiment with air beam paths.
dc.languageeng
dc.publisherElsevier Science
dc.relationinfo:eu-repo/semantics/altIdentifier/url/https://www.sciencedirect.com/science/article/pii/S0168900218303723
dc.relationinfo:eu-repo/semantics/altIdentifier/doi/https://doi.org/10.1016/j.nima.2018.03.036
dc.rightshttps://creativecommons.org/licenses/by-nc-sa/2.5/ar/
dc.rightsinfo:eu-repo/semantics/restrictedAccess
dc.subjectBACKGROUND RADIATION
dc.subjectINELASTIC X-RAY SCATTERING
dc.subjectJOHANN SPECTROMETER
dc.subjectX-RAY EMISSION SPECTROSCOPY
dc.titleBackground radiation in inelastic X-ray scattering and X-ray emission spectroscopy. A study for Johann-type spectrometers
dc.typeinfo:eu-repo/semantics/article