dc.description.abstract | En este trabajo se muestran las modificaciones realizadas al actual CPAP de flujo variable de válvula Benveniste que está siendo utilizado en varias clínicas de la ciudad de Medellín.
En la primera modificación se le realizaron mejoras a la seguridad del paciente en el momento de ser aplicada la terapia, por medio del diseño y construcción de un monitor de presión, el cual se encarga de alarmar al personal de salud, si hay cambios en la presión del flujo de aire a la salida de la válvula Benveniste en un rango de + 1 cm de agua. Antes de hacer esta modificación, para asegurarse que la presión estuviera en el nivel deseado, era necesario desconectar al paciente de la terapia e introducir la manguera que lleva el flujo de aire-oxígeno en un recipiente con agua, donde se miraba el valor de la presión en centímetros de agua.
También se modificó la vía de administración del flujo de aire-oxígeno al paciente durante el tiempo que se encuentre en terapia; el dispositivo existente solo contiene un prong nasal, que en realidad es una parte de un tubo endotraqueal; se realizó el diseño e implementación de prongs binasales al dispositivo. | |
dc.description.abstract | ABSTRACT:
This work involves some modifications to the current variable flow CPAP Benveniste valve being used in some clinics in the city of Medellin.
In the first modification were made improvements to patient safety at the time of therapy to be applied, through the design and construction of a pressure monitor, which is responsible for alarming the health personnel, if there are changes in air flow pressure at the end of the Benveniste valve in a range of ± 1 cm of water. Before this change, to ensure that the pressure was at the desired level, it was necessary to disconnect the patient from therapy and insert the hose that carries air-oxygen flow in a water container, where the value of pressure was looked in centimeters of water.
Also it was changed the route of administration of air-oxygen flow to the patient during the time of therapy, the only existing device contains a mononasal prong, which is a part of an endotracheal tube; the design and implementation of binasales prongs to the device were performed | |
dc.language | spa | |
dc.publisher | Universidad EIA | |
dc.publisher | Ingeniería Biomédica | |
dc.publisher | Escuela de Ciencias de la Vida | |
dc.publisher | Envigado, Antioquia, Colombia | |
dc.rights | https://creativecommons.org/licenses/by-nc-nd/4.0/ | |
dc.rights | info:eu-repo/semantics/openAccess | |
dc.rights | Atribución-NoComercial-SinDerivadas 4.0 Internacional (CC BY-NC-ND 4.0) | |
dc.rights | Derechos Reservados - Universidad EIA, 2010 | |
dc.title | Adaptaciones para el sistema de CPAP Nasal Benveniste | |
dc.type | Trabajo de grado - Pregrado | |