Artículos de revistas
Evaluation of three different osteosynthesis methods for mandibular angle fractures: Vertical load test
Fecha
2016-01-01Registro en:
Journal of Craniofacial Surgery, v. 27, n. 7, p. 1770-1773, 2016.
1536-3732
1049-2275
10.1097/SCS.0000000000003076
2-s2.0-85028272634
8492596401380580
Autor
Universidade Estadual Paulista (Unesp)
Institución
Resumen
The aim of this study was to evaluate the mechanical behavior of different rigid fixation methods in mandible angle fractures. Three different plates were tested: one 4-hole grid miniplate, one 8-hole curved grid miniplate, and one 4-hole straight miniplate. For the loading tests, 30 polyurethane hemimandibles sectioned at the angle area to simulate a simple angle fracture were used: 10 hemimandibles for each group. The mechanical evaluation revealed that the straight plate positioned at the tension zone presented the best results in supporting vertical loading in the predefined displacement moments of 1, 3, and 5 mm. Regarding the grid plates, there was no statistically significant difference between the smaller and the larger plate. According to this sample, the straight plate positioned in the tension zone presented better results than the other 2 different grid plates positioned at the neutral zone.