Occupational radiation exposure in vascular interventional radiology: A complete evaluation of different body regions
Date
2016-08-01Registration in:
Physica Medica, v. 32, n. 8, p. 1019-1024, 2016.
1724-191X
1120-1797
10.1016/j.ejmp.2016.06.014
2-s2.0-84991206833
2-s2.0-84991206833.pdf
Author
Universidade Estadual Paulista (Unesp)
Sapra Landauer
Universidade de São Paulo (USP)
Institutions
Abstract
Purpose To perform a complete evaluation on radiation doses, received by primary and assistant medical staff, while performing different vascular interventional radiology procedures. Materials and methods We evaluated dose received in different body regions during three categories of vascular procedures: lower limb angiography (Angiography), lower limb percutaneous transluminal angioplasty (Angioplasty) and stent graft placement for abdominal aortic aneurysm treatment (A. A. A. Treatment). We positioned the dosimeters near the eye lens, thyroid, chest, abdomen, hands, and feet of the interventional physicians. Equivalent dose was compared with annual dose limits for workers in order to determine the maximum number of procedures per year that each physician could perform. We assessed 90 procedures. Results We found the highest equivalent doses in the A. A. A. Treatment, in which 90% of the evaluations indicated at least one region receiving more than 1 mSv per procedure. Angioplasty was the only procedural modality that provided statistically different doses for different professionals, which is an important aspect on regards to radiological protection strategies. In comparison with the dose limits, the most critical region in all procedures was the eye lens. Conclusions Since each body region of the interventionist is exposed to different radiation levels, dose distribution measurements are essential for radiological protection strategies. These results indicate that dosimeters placed in abdomen instead of chest may represent more accurately the whole body doses received by the medical staff. Additional dosimeters and a stationary shield for the eye lens are strongly recommended.