Preventing Chaotic Motion in Tapping-Mode Atomic Force Microscope
Fecha
2014-01-01Registro en:
Journal of Control, Automation and Electrical Systems, v. 25, n. 6, p. 732-740, 2014.
2195-3899
2195-3880
10.1007/s40313-014-0144-4
2-s2.0-84919913410
2-s2.0-84919913410.pdf
Autor
Universidade de São Paulo (USP)
Universidade Estadual Paulista (Unesp)
Institución
Resumen
During the last 30 years the Atomic Force Microscopy became the most powerful tool for surface probing in atomic scale. The Tapping-Mode Atomic Force Microscope is used to generate high quality accurate images of the samples surface. However, in this mode of operation the microcantilever frequently presents chaotic motion due to the nonlinear characteristics of the tip-sample forces interactions, degrading the image quality. This kind of irregular motion must be avoided by the control system. In this work, the tip-sample interaction is modelled considering the Lennard-Jones potentials and the two-term Galerkin aproximation. Additionally, the State Dependent Ricatti Equation and Time-Delayed Feedback Control techniques are used in order to force the Tapping-Mode Atomic Force Microscope system motion to a periodic orbit, preventing the microcantilever chaotic motion.