dc.description.abstract | Foi implementado um instrumento com a finalidade de medir a intensidade (taxa de fluência de energia) de fótons na faixa de mamografia (22 a 33 kVp) e a tensão do tubo de um mamógrafo. O instrumento é constituído por uma câmara com sensor piroelétrico de metaniobato de chumbo (PbNb2¬O¬¬6) e seu circuito de condicionamento de sinais. Utilizou-se poliácido lático para confeccionar a câmara piroelétrica em uma impressora 3D. O circuito de condicionamento de sinais é um conversor corrente-tensão de alta sensibilidade. A intensidade da radiação foi determinada a partir da taxa de exposição medida com uma câmara de ionização. Uma simulação do espectro de emissão do molibdênio, na faixa de 22 a 33 kVp, foi realizada com o sensor piroelétrico e o software DTSA-II, e comparada com a o feita com um sensor de Si(Li), comumente empregado. Os resultados obtidos mostram que a resposta do instrumento implementado é linear com a tensão do tubo de raios-X e com a intensidade da radiação, e que o espectro de emissão do molibdênio pode ser determinado com o instrumento piroelétrico. | |
dc.description.abstract | An instrument to measure the intensity (energy fluence rate) of photons in the mammography range (22 to 33 kVp) and the tube voltage of a mammograph. The instrument consists of a chamber with a pyroelectric sensor of lead methaniobate (PbNb2O6) and its signal conditioning circuit. Polyacid was used to make the pyroelectric chamber in a 3D printer. The signal conditioning circuit is a high sensitivity current-voltage converter. The intensity of the radiation was determined from the exposure rate measured with an ionization chamber. A simulation of the emission spectrum for molybdenum, in the range of 22 to 33 kVp, was performed with the pyroelectric sensor and the DTSA-II software, and compared to that simulated using the commonly used Si (Li) sensor. The results show that the response of the implemented instrument is linear with the X-ray tube voltage and the intensity of the radiation, and that the emission spectrum of molybdenum can be determined with a pyroelectric instrument. | |
dc.language | por | |
dc.publisher | Universidade Estadual Paulista (Unesp) | |
dc.rights | Acesso aberto | |
dc.subject | Sensor piroelétrico | |
dc.subject | Mamografia | |
dc.subject | Taxa de fluência de energia | |
dc.subject | Raios-X | |
dc.subject | Dosimetria | |
dc.subject | Simulação de raios-X | |
dc.subject | Pyroelectric sensor | |
dc.subject | Mammography | |
dc.subject | Energy fluency rate | |
dc.subject | X ray | |
dc.subject | Dosimetry | |
dc.subject | Simulation of X-rays | |
dc.title | Implementação de um instrumento para medição da intensidade de fótons na faixa da mamografia | |
dc.type | Tesis | |