In vitro evaluation of conventional and locking miniplate/screw systems for the treatment of mandibular angle fractures
Fecha
2010-11-01Registro en:
International Journal of Oral and Maxillofacial Surgery. Edinburgh: Churchill Livingstone, v. 39, n. 11, p. 1109-1114, 2010.
0901-5027
10.1016/j.ijom.2010.06.019
WOS:000284386000011
5535418670745125
Autor
Universidade do Sagrado Coração (USC)
Universidade Estadual Paulista (Unesp)
Tufts Univ
Institución
Resumen
This in vitro study evaluated the influence of the type of miniplate and the number of screws installed in the proximal and distal segments on the stability and resistance of Champy's osteosynthesis in mandibular angle fractures. Sixty polyurethane hemimandibles with bone-like consistency were randomly assigned to four groups (n = 15) and sectioned in the mandibular angle region to simulate fracture. The bone segments were fixed by different osteosynthesis methods using 2.0 mm miniplates and 2.0 mm x 6 mm rnonocortical screws. In groups 1 and 2, two conventional (G1) or locking (G2) screws were installed in each bone segment using a conventional (G1) or a locking (02) straight miniplate; in groups 3 and 4, three conventional (03) or locking (04) screws were installed in the proximal segment and four conventional (G3) or locking (04) screws were installed in the distal segment using a conventional (G3) or a locking (G4) seven-hole straight miniplate. The hemimandibles were loaded in compressive strength until a 4 mm displacement occurred between the segments, vertically or horizontally. Locking plate/screw systems provided significantly greater resistance to displacement than conventional ones (p < .01). Locking miniplates offered more resistance than conventional miniplates. Long locking miniplates provided greater stability than short ones.