dc.identifierGarcia Inza, Mariano Andrés; Carbonetto, Sebastián Horacio; Lipovetzky, José; Faigon, Adrian Nestor; Radiation sensor based on MOSFETs mismatch amplification for radiotherapy applications; Institute of Electrical and Electronics Engineers; Ieee Transactions on Nuclear Science; 63; 3; 6-2016; 1784-1789
dc.identifier0018-9499
dc.identifierhttp://hdl.handle.net/11336/20123
dc.identifier1558-1578
dc.identifierCONICET Digital
dc.identifierCONICET
dc.description.abstractIn this paper we present a new dosimeter based on a pair of thick gate oxide MOSFET sensors. A differential circuit topology with a feedback loop provides stabilized output with selectable sensitivity amplification for real time in vivo dosimetry. Radiation response shows a wide linear output range and an effective thermal rejection. These properties make this circuit suitable for dose control in radiotherapy applications.
dc.languageeng
dc.publisherInstitute of Electrical and Electronics Engineers
dc.relationinfo:eu-repo/semantics/altIdentifier/url/http://ieeexplore.ieee.org/document/7464334/
dc.relationinfo:eu-repo/semantics/altIdentifier/doi/http://dx.doi.org/10.1109/TNS.2016.2560172
dc.rightshttps://creativecommons.org/licenses/by-nc-sa/2.5/ar/
dc.rightsinfo:eu-repo/semantics/openAccess
dc.subjectMosfet Dosimeter
dc.subjectRadiotherapy
dc.subjectIonizing Radiation
dc.subjectSolid-State Detectors
dc.titleRadiation sensor based on MOSFETs mismatch amplification for radiotherapy applications
dc.typeinfo:eu-repo/semantics/article
dc.typeinfo:ar-repo/semantics/artículo
dc.typeinfo:eu-repo/semantics/publishedVersion


Este ítem pertenece a la siguiente institución