dc.identifierJerez, Javier; Isaguirre, Andrea Celeste; Bazan, Cristian Roberto; Martinez, Luis Dante; Cerutti, Estela Soledad; Determination of scandium in acid mine drainage by ICP-OES with flow injection on-line preconcentration using oxidized multiwalled carbon nanotubes; Elsevier; Talanta; 124; 11-2014; 89-94
dc.identifier0039-9140
dc.identifierhttp://hdl.handle.net/11336/5765
dc.identifier.urihttp://repositorioslatinoamericanos.uchile.cl/handle/2250/1885443
dc.description.abstractAn on-line scandium preconcentration and determination system implemented with inductively coupled plasma optical emission spectrometry associated with flow injection was studied. Trace amounts of scandium were preconcentrated by sorption on a minicolumn packed with oxidized multiwalled carbon nanotubes, at pH 1.5. The retained analyte was removed from the minicolumn with 30% (v/v) nitric acid. A total enrichment factor of 225-fold was obtained within a preconcentration time of 300 s (for a 25 mL sample volume). The overall time required for preconcentration and elution of 25 mL of sample was about 6 min; the throughput was about 10 samples per hour. The value of the detection limit was 4 ng L1 and the precision for 10 replicate determinations at 100 ng L1 Sc level was 5% relative standard deviation, calculated from the peak heights obtained. The calibration graph using the preconcentration system was linear with a correlation coefficient of 0.9996 at levels near the detection limits up to at least 10 mg L1 . After optimization, the method was successfully applied to the determination of Sc in an acid drainage from an abandoned mine located in the province of San Luis, Argentina
dc.languageeng
dc.publisherElsevier
dc.relationinfo:eu-repo/semantics/altIdentifier/url/http://www.sciencedirect.com/science/article/pii/S003991401400126X
dc.relationinfo:eu-repo/semantics/altIdentifier/doi/10.1016/j.talanta.2014.02.028
dc.relationinfo:eu-repo/semantics/altIdentifier/doi/http://dx.doi.org/10.1016/j.talanta.2014.02.028
dc.rightshttps://creativecommons.org/licenses/by-nc-nd/2.5/ar/
dc.rightsinfo:eu-repo/semantics/restrictedAccess
dc.subjectScandium
dc.subjectAcid mine drainage
dc.subjectOx-CNTs
dc.subjectICP-OES
dc.titleDetermination of scandium in acid mine drainage by ICP-OES with flow injection on-line preconcentration using oxidized multiwalled carbon nanotubes
dc.typeArtículos de revistas
dc.typeArtículos de revistas
dc.typeArtículos de revistas


Este ítem pertenece a la siguiente institución