dc.creatorJuan de Paz, Leornardo
dc.creatorGraf, Daniel Adolfo
dc.creatorScandizzi, Angel Luis
dc.creatorGuibert, Edgardo Elvio
dc.creatorRodriguez, Joaquin Valentin
dc.date.accessioned2017-08-07T18:50:27Z
dc.date.accessioned2018-11-06T14:16:07Z
dc.date.available2017-08-07T18:50:27Z
dc.date.available2018-11-06T14:16:07Z
dc.date.created2017-08-07T18:50:27Z
dc.date.issued2013-08
dc.identifierJuan de Paz, Leornardo; Graf, Daniel Adolfo; Scandizzi, Angel Luis; Guibert, Edgardo Elvio; Rodriguez, Joaquin Valentin; A device to record ultra-rapid cooling profiles; Cryo Letters; Cryo-letters; 34; 4; 8-2013; 360-368
dc.identifier0143-2044
dc.identifierhttp://hdl.handle.net/11336/21966
dc.identifierCONICET Digital
dc.identifierCONICET
dc.identifier.urihttp://repositorioslatinoamericanos.uchile.cl/handle/2250/1884587
dc.description.abstractThis work deals with the construction and performance of a measuring system capable of estimating temperature at sufficiently high speed (up to 1000 samples/sec). Due to its simple design and the utilization of standard materials, it could serve to recording the cooling profile of ultra-rapid procedures. An immersion device was also developed with the purpose of normalize the penetration speed of the sample in the LN2. The device allows also the comparative analysis of different cooling profiles. The system consists of an immersion device of the sample in the cooling agent, a temperature measurement system developed by Kleihans F and a laptop computer. To test the system, we recorded the cooling profiles of 10 L of distilled water and 6 M glycerol solution, obtaining a cooling rate of 8732 °C.min-1 and 4441 °C.min-1 respectively. Also we determine a cooling rate of 204.012 °C.min-1 during the immersion of the thermocouple assembly in LN2. Although, the same device, with small technical modifications related to the handling of the sample, could be used to evaluate the recovery from LN2 temperature to room temperature (re-warming).
dc.languageeng
dc.publisherCryo Letters
dc.relationinfo:eu-repo/semantics/altIdentifier/url/http://www.cryoletters.org/Abstracts/vol_34_4_2013.htm#360
dc.rightshttps://creativecommons.org/licenses/by-nc-sa/2.5/ar/
dc.rightsinfo:eu-repo/semantics/restrictedAccess
dc.subjectCOOLING
dc.subjectVITRIFICATION
dc.titleA device to record ultra-rapid cooling profiles