Enhanced high-performance liquid chromatography method for the determination of retinoic acid in plasma: Development, optimization and validation
Fecha
2014-06Registro en:
Goicoechea, Hector Casimiro; Martínez Galera, María; Gil García, María D.; Teglia, Carla Mariela; Enhanced high-performance liquid chromatography method for the determination of retinoic acid in plasma: Development, optimization and validation; Elsevier; Journal of Chromatography - A; 1353; 6-2014; 40-48
0021-9673
CONICET Digital
CONICET
Autor
Teglia, Carla Mariela
Gil García, María D.
Martínez Galera, María
Goicoechea, Hector Casimiro
Resumen
When determining endogenous compounds in biological samples, the lack of blank or analyte-free matrix samples involves the use of alternative strategies for calibration and quantitation. This article deals with the development, optimization and validation of a high performance liquid chromatography method for the determination of retinoic acid in plasma, obtaining at the same time information about its isomers, taking into account the basal concentration of these endobiotica. An experimental design was used for the optimization of three variables: mobile phase composition, flow rate and column temperature through a central composite design. Four responses were selected for optimization purposes (area under the peaks, quantity of peaks, analysis time and resolution between the first principal peak and the following one). The optimum conditions resulted in a mobile phase consisting of methanol 83.4% (v/v), acetonitrile 0.6% (v/v) and acid aqueous solution 16.0% (v/v); flow rate of 0.68 mL min−1 and an column temperature of 37.10 °C. Detection was performed at 350 nm by a diode array detector. The method was validated following a holistic approach that included not only the classical parameters related to method performance but also the robustness and the expected proportion of acceptable results lying inside predefined acceptability intervals, i.e., the uncertainty of measurements. The method validation results indicated a high selectivity and good precision characteristics that were studied at four concentration levels, with RSD less than 5.0% for retinoic acid (less than 7.5% for the LOQ concentration level), in intra and inter-assay precision studies. Linearity was proved for a range from 0.00489 to 15.109 ng mL−1 of retinoic acid and the recovery, which was studied at four different fortification levels in phuman plasma samples, varied from 99.5% to 106.5% for retinoic acid. The applicability of the method was demonstrated by determining retinoic acid and obtaining information about its isomers in human and frog plasma samples from different origins.