dc.creatorPACHECO, Bruno D.
dc.creatorVALERIO, Jaqueline
dc.creatorANGNES, Lucio
dc.creatorPEDROTTI, Jairo J.
dc.date.accessioned2012-10-20T05:23:32Z
dc.date.accessioned2018-07-04T15:49:22Z
dc.date.available2012-10-20T05:23:32Z
dc.date.available2018-07-04T15:49:22Z
dc.date.created2012-10-20T05:23:32Z
dc.date.issued2011
dc.identifierANALYTICA CHIMICA ACTA, v.696, n.1/Fev, p.53-58, 2011
dc.identifier0003-2670
dc.identifierhttp://producao.usp.br/handle/BDPI/31233
dc.identifier10.1016/j.aca.2011.04.004
dc.identifierhttp://dx.doi.org/10.1016/j.aca.2011.04.004
dc.identifier.urihttp://repositorioslatinoamericanos.uchile.cl/handle/2250/1627871
dc.description.abstractA fast and robust analytical method for amperometric determination of hydrogen peroxide (H(2)O(2)) based on batch injection analysis (BIA) on an array of gold microelectrodes modified with platinum is proposed. The gold microelectrode array (n = 14) was obtained from electronic chips developed for surface mounted device technology (SMD), whose size offers advantages to adapt them in batch cells. The effect of the dispensing rate, volume injected, distance between the platinum microelectrodes and the pipette tip, as well as the volume of solution in the cell on the analytical response were evaluated. The method allows the H(2)O(2) amperometric determination in the concentration range from 0.8 mu mol L(-1) to 100 mu mol L(-1). The analytical frequency can attain 300 determinations per hour and the detection limit was estimated in 0.34 mu mol L(-1) (3 sigma). The anodic current peaks obtained after a series of 23 successive injections of 50 mu L of 25 mu mol L(-1) H(2)O(2) showed an RSD < 0.9%. To ensure the good selectivity to detect H(2)O(2), its determination was performed in a differential mode, with selective destruction of the H(2)O(2) with catalase in 10 mmol L(-1) phosphate buffer solution. Practical application of the analytical procedure involved H(2)O(2) determination in rainwater of Sao Paulo City. A comparison of the results obtained by the proposed ampermetric method with another one which combines flow injection analysis (FIA) with spectrophotometric detection showed good agreement. (C) 2011 Elsevier B.V. All rights reserved.
dc.languageeng
dc.publisherELSEVIER SCIENCE BV
dc.relationAnalytica Chimica Acta
dc.rightsCopyright ELSEVIER SCIENCE BV
dc.rightsrestrictedAccess
dc.subjectHydrogen peroxide
dc.subjectAmperometry
dc.subjectBatch injection analysis
dc.subjectGold microelectrode arrays
dc.subjectRainwater
dc.titleFast batch injection analysis of H(2)O(2) using an array of Pt-modified gold microelectrodes obtained from split electronic chips
dc.typeArtículos de revistas


Este ítem pertenece a la siguiente institución