dc.creator | COSTA, Alessandro M. | |
dc.creator | BADIN, Romulo S. | |
dc.creator | LEITE, Marina S. | |
dc.creator | CALDAS, Linda V. E. | |
dc.date.accessioned | 2012-10-19T14:19:07Z | |
dc.date.accessioned | 2018-07-04T15:02:36Z | |
dc.date.available | 2012-10-19T14:19:07Z | |
dc.date.available | 2018-07-04T15:02:36Z | |
dc.date.created | 2012-10-19T14:19:07Z | |
dc.date.issued | 2008 | |
dc.identifier | APPLIED RADIATION AND ISOTOPES, v.66, n.10, p.1403-1405, 2008 | |
dc.identifier | 0969-8043 | |
dc.identifier | http://producao.usp.br/handle/BDPI/21018 | |
dc.identifier | 10.1016/j.apradiso.2008.03.007 | |
dc.identifier | http://dx.doi.org/10.1016/j.apradiso.2008.03.007 | |
dc.identifier.uri | http://repositorioslatinoamericanos.uchile.cl/handle/2250/1617796 | |
dc.description.abstract | A tandem ionization chamber was developed for quality control programs of X-ray equipment used in conventional radiography and mammography. A methodology for the use of the tandem chamber in the constancy check of diagnostic X-ray beam qualities was established. The application at a medical X-ray imaging facility of this established methodology is presented. The use of the tandem chamber in the constancy check of diagnostic X-ray beam qualities is a useful method to control the performance of the X-ray equipment. (c) 2008 Elsevier Ltd. All rights reserved. | |
dc.language | eng | |
dc.publisher | PERGAMON-ELSEVIER SCIENCE LTD | |
dc.relation | Applied Radiation and Isotopes | |
dc.rights | Copyright PERGAMON-ELSEVIER SCIENCE LTD | |
dc.rights | restrictedAccess | |
dc.subject | ionization chambers | |
dc.subject | X-radiation | |
dc.subject | quality control | |
dc.subject | diagnostic radiology | |
dc.subject | tandem system | |
dc.title | Constancy check of beam quality in conventional diagnostic X-ray equipment | |
dc.type | Artículos de revistas | |