dc.description.abstractNeste trabalho foram estudadas as propriedades mecânicas de um novo sistema de osteossíntese metálica (SPS® - Sistema Pengo de Síntese) por meio de ensaios mecânicos de flexão e torção. Os SPS® foram montados e fixados em bastões cilíndricos de madeira. Para os ensaios mecânicos foram adotados nove grupos de modelos em função dos três comprimentos de placas e das três distâncias de montagem entre as placas. Os ensaios de flexão foram realizados com o SPS® na posição lateral em relação ao eixo de movimentação da máquina de ensaios. Para os ensaios de torção não se adotou nenhuma posição específica do SPS®. Os resultados das propriedades de rigidez e da deflexão do SPS®, nos ensaios mecânicos de flexão e torção, tiveram a interferência das variantes números de furos nas placas e o espaço entre as placas, ou seja, quanto maior o número de furos nas placas e menor o espaço entre elas maior a rigidez e menor a deflexão. Conclui-se que a rigidez e a deflexão variaram proporcionalmente ao número de furos nas placas e ao espaço entre elas.
dc.description.abstractIn this study, the mechanical properties of a new metallic system for bone synthesis (SPS® - Synthesis Pengo System) were investigated by means of flexion and torsion mechanical tests. The SPS® systems were assembled in cylindrical wood sticks using stainless steel cortical bolts. Nine groups of models were adopted for tests according to the three plate lengths and the three assembly distances. Flexion tests were performed with the SPS® systems positioned laterally to the assay machine's dislocation axis. For the torsion tests, the SPS® systems were placed in no specific position. The SPS® stiffness and deflection results obtained with the flexion and torsion mechanical tests were dependent on the number of holes on the plates and on the distance between the plates mounted in the wood stick, that is, the higher the number of holes and the shorter the distance between plates, the higher the stiffness and the lower the deflection. It can be concluded that stiffness and deflection changed proportionally to the number of the holes in the plates and the distance between them.
dc.languagepor
dc.publisherAtha Comunicação & Editora
dc.relationActa Ortopédica Brasileira
dc.rightsCopyright Atha Comunicação & Editora
dc.rightsopenAccess
dc.subjectPlacas ósseas
dc.subjectFixação de fratura
dc.subjectFixação interna de fraturas
dc.subjectBone plates
dc.subjectFracture fixation
dc.subjectFracture fixation internal
dc.titlePropriedades mecânicas de um sistema de osteossíntese de estabilidade relativa