Self-reinforced polymeric plates and screws for rigid internal fixation of fractures

dc.creatorMorita, Augusto Teruo
dc.date2006
dc.date2006-03-15T00:00:00Z
dc.date2017-03-28T23:37:31Z
dc.date2017-10-16T13:03:33Z
dc.date2017-03-28T23:37:31Z
dc.date2017-10-16T13:03:33Z
dc.date.accessioned2018-03-29T05:38:38Z
dc.date.available2018-03-29T05:38:38Z
dc.identifier(Broch.)
dc.identifierMORITA, Augusto Teruo. Placas e parafusos polimericos auto-reforçados para fixação interna rigida de fraturas. 2006. 156p. Tese (doutorado) - Universidade Estadual de Campinas, Instituto de Quimica, Campinas, SP. Disponível em: <http://www.bibliotecadigital.unicamp.br/document/?code=vtls000383823>. Acesso em: 28 mar. 2017.
dc.identifierhttp://repositorio.unicamp.br/jspui/handle/REPOSIP/248658
dc.identifier.urihttp://repositorioslatinoamericanos.uchile.cl/handle/2250/1361052
dc.descriptionOrientador: Marco-Aurelio De Paoli
dc.descriptionTese (doutorado) - Universidade Estadual de Campinas, Instituto de Quimica
dc.descriptionResumo: O objetivo deste trabalho foi preparar placas e parafusos auto-reforçados a partir do copolímero poli(L/DL ácido láctico), para serem utilizados na fixação interna rígida de fraturas. Este copolímero pode ser hidrolisado quando inserido no corpo humano e seus produtos de degradação são expelidos através do ciclo metabólico. Foi construído um aparelho de autoreforçamento que deve ser acoplado a uma Máquina Universal de Ensaios, capaz de orientar uniaxialmente amostras poliméricas através da passagem por uma matriz aquecida de dimensões menores que a amostra. Foram realizados ensaios prévios de auto-reforçamento com amostras de poliestireno. Ensaios de auto-reforçamento com o copolímero poli(L/DL ácido láctico) mostraram que o auto-reforçamento induziu um aumento nas propriedades mecânicas, medidas sob flexão, de ~60% no módulo de elasticidade e de ~90% na tensão máxima. A amostra moldada apresentou linhas brancas quando dobrada a temperatura ambiente, indicando a fragilização. Por outro lado, as amostras auto-reforçadas podem ser dobradas a temperatura ambiente sem que ocorresse ruptura frágil ou fragilização. Através de ensaios de difração de raios-X determinou-se que a amostra é semi-cristalina após o auto-reforçamento, apesar dos ensaios de calorimetria diferencial de varredura não identificarem positivamente a formação desta estrutura semi-cristalina. A esterilização por raios-g causa uma grande queda na massa molar, porém não altera as propriedades mecânicas. Ensaios de degradação in-vitro, com solução tampão de fosfato numa temperatura de 37 °C, das amostras esterilizadas mostraram que a massa molar sofre uma queda durante todo o ensaio de degradação e as propriedades mecânicas sofrem uma pequena queda após 15 semanas. As amostras auto-reforçadas podem ser usinadas na forma de placas e parafusos. As placas usinadas do copolímero de poli(ácido láctico) ainda podem ser dobradas a temperatura ambiente sem que ocorra quebra ou formação de linhas brancas
dc.descriptionAbstract: In this work self-reinforced plates and screws to be used for rigid internal fixation of fractures were produced using the poly(L/DL-lactic acid) copolymer. This copolymer can be hydrolyzed when implanted in the human body and its degradation products are expelled by the metabolic cycle. A self-reinforcement module was built. It can be attached to a Universal Mechanical Testing Machine, which is capable to uni-axially orient the polymeric sample though a heated die. Initial tests of processing and self-reinforcement were conducted with polystyrene. Self-reinforcement tests were conducted with the poly(L/DL-lactic acid) copolymer. The results show that the die-drawing process induced an increase in the flexural mechanical properties of ~60% in the elastic modulus and of ~90% in the maximum tension, both measured under flexion conditions. White lines (crazing) appears when the molded sample was bent at ambient temperature, whereas the self-reinforced sample could be bent in the same conditions without fracture or sample fragilization. X-ray diffraction results indicated that the self-reinforced sample exhibits a semi-crystalline structure; notwithstanding the differential scanning calorimetry results do not conclusively identify the sample morphology. g-rays sterilization caused a decrease in the molar mass, but the mechanical properties were not affected. In-vitro degradation test were conducted at 37 °C in a buffered aqueous solution. Sterilized samples show a larger decrease in the molar mass since the beginning of the test, but the mechanical properties exhibits a smaller decrease during the test. The self-reinforced samples can be molded to the form of plates and screws. The final sample could be bent at ambient temperature without fragilization
dc.descriptionDoutorado
dc.descriptionQuimica Inorganica
dc.descriptionDoutor em Ciencias
dc.format156p. : il.
dc.formatapplication/pdf
dc.languagePortuguês
dc.publisher[s.n.]
dc.subjectPolímeros auto-reforçados
dc.subjectHips
dc.subjectNoryl
dc.subjectAntiestaticos
dc.subjectFraturas - Fixação interna
dc.subjectMateriais biodegradaveis
dc.subjectPoli (ácido láctico)
dc.subjectSelf-reinforced polymers
dc.subjectInternal fracture fixation
dc.subjectBiodegradable materials
dc.subjectPoly (lactic acid)
dc.titlePlacas e parafusos polimericos auto-reforçados para fixação interna rigida de fraturas
dc.titleSelf-reinforced polymeric plates and screws for rigid internal fixation of fractures
dc.typeTesis