Mechanical and microstructural evaluation of two internal fixation systems used in mandibular sagittal split ramus osteotomy

dc.creatorMendes, Marcelo Breno Meneses, 1984-
dc.date2014
dc.date2017-04-02T00:17:32Z
dc.date2017-07-12T19:20:48Z
dc.date2017-04-02T00:17:32Z
dc.date2017-07-12T19:20:48Z
dc.date.accessioned2018-03-29T03:41:12Z
dc.date.available2018-03-29T03:41:12Z
dc.identifierMENDES, Marcelo Breno Meneses. Estudo mecânico e microestrutural em dois sistemas de fixação interna usados em osteotomia sagital mandibular. 2014. 73 f. Tese (doutorado) - Universidade Estadual de Campinas, Faculdade de Odontologia de Piracicaba, Piracicaba, SP. Disponível em: <http://www.bibliotecadigital.unicamp.br/document/?code=000920099>. Acesso em: 1 abr. 2017.
dc.identifierhttp://repositorio.unicamp.br/jspui/handle/REPOSIP/288763
dc.identifier.urihttp://repositorioslatinoamericanos.uchile.cl/handle/2250/1334806
dc.descriptionOrientador: Roger William Fernandes Moreira
dc.descriptionTese (doutorado) - Universidade Estadual de Campinas, Faculdade de Odontologia de Piracicaba
dc.descriptionResumo: O objetivo deste estudo foi avaliar, através da análise de rugosidade superficial, microscopia eletrônica de varredura (MEV), análise de composição através de espectroscopia de raios-X e análise de resistência mecânica, dois diferentes sistemas de fixação interna usados em fixação de osteotomia sagital do ramo mandibular. Foram selecionadas cinco placas retas do sistema 2,0 mm com quatro furos e 20 parafusos de 6 mm de comprimento de dois sistemas de fixação interna, grupo I: Tóride (Tóride Indústria e Comércio Ltda. -Mogi Mirim, São Paulo -Brasil) e grupo II: Medartis (Medartis AG -Basel, Suíça). Para os testes mecânicos, foram selecionadas dez hemimandíbulas de poliuretano, dividindo-as igualmente entre os grupos I e II. Inicialmente, foi realizada análise de rugosidade superficial em todas as placas dos dois sistemas, e os dados foram submetidos à análise de variância e teste de Tukey (p>0,05) para a comparação entre as médias. Em seguida, as mesmas placas foram avaliadas no espectrômetro por meio de difração de raios-X e logo após foram analisadas no MEV. Por último, foi realizado o teste mecânico, no qual as hemimandíbulas de poliuretano foram fixadas e submetidas ao teste de carregamento (1mm/min). Os dados foram analisados quanto à carga de pico (em Newtons) e o deslocamento de pico (em mm) para todos os grupos estudados, e posteriormente submetidos ao teste t. Os grupos testados apresentaram vestígios superficiais dos processos de usinagem, principalmente nas margens dos elos. O grupo II apresentou revestimento com uma camada de nitreto de titânio. A análise da rugosidade revelou diferença entre as regiões de centro e margem da placa para o grupo I; diferente do grupo II, o qual apresentou valor de rugosidade menor e mais uniforme. O teste mecânico revelou que o grupo II apresentou maior carga de pico. Os dados demonstraram que há diferença microestrutural e mecânica entre os grupos testados.
dc.descriptionAbstract: The aim of this study was to evaluate, through surface roughness analysis, scanning electron microscopy (SEM), composition analysis by X-ray spectroscopy and analysis of mechanical strength, two different systems used in internal fixation of mandibular sagittal split osteotomy fixation. Five straight 2.0 mm system plates with four holes and 20 screws 6 mm length of two internal fixation systems were selected, group I: Toride (Toride Indústria e Comércio Ltda -Mogi Mirim, São Paulo -Brazil.) and group II: Medartis (Medartis AG -Basel, Switzerland). For mechanical testing, ten polyurethane hemimandibles were used, distributing them equally into groups I and II. Initially, surface roughness analysis was performed on all plates of the two systems, and the data were subjected to analysis of variance and Tukey's test (p> 0.05) for comparison between means. Then, the same plates were evaluated in the spectrometer by means of X-ray diffraction and after this evaluation, they were subjected to SEM. Finally, mechanical testing was performed, in which the polyurethane mandibles were fixed and subjected to the load (1mm/min) test on an Instron testing machine, Model 4411 (Instron Corp., Norwood, MA). Data were analyzed for peak load (in Newtons) and peak displacement (in mm) for all groups, and then subjected to t test. The groups tested showed superficial traces of machining processes, especially on links periphery. Group II had coated with a layer of titanium nitride. The roughness analysis revealed differences among the regions of the center and edge of the plate for group I; group II showed roughness value smaller and more uniform, than group I.The mechanical test showed that group II had higher peak load. The data demonstrated that there are microstructural and mechanical differences between the tested groups.
dc.descriptionDoutorado
dc.descriptionCirurgia e Traumatologia Buco-Maxilo-Faciais
dc.descriptionDoutor em Clínica Odontológica
dc.format73 f. : il.
dc.formatapplication/pdf
dc.languagePortuguês
dc.publisher[s.n.]
dc.subjectPropriedades de superficie
dc.subjectEspectrometria por raios X
dc.subjectFraturas - Fixação interna
dc.subjectTécnicas de química analítica
dc.subjectTitânio
dc.subjectSurface properties
dc.subjectSpectrometry, X-ray emission
dc.subjectFracture fixation, internal
dc.subjectChemistry techniques, Analytical
dc.subjectTitanium
dc.titleEstudo mecânico e microestrutural em dois sistemas de fixação interna usados em osteotomia sagital mandibular
dc.titleMechanical and microstructural evaluation of two internal fixation systems used in mandibular sagittal split ramus osteotomy
dc.typeTesis


Este ítem pertenece a la siguiente institución