dc.creatorBassani, Mariana A
dc.creatorMezzacappa Filho, Francisco
dc.creatorCoppo, Maria Regina C
dc.creatorMarba, Sérgio T M
dc.date
dc.date2015-11-27T13:15:01Z
dc.date2015-11-27T13:15:01Z
dc.date.accessioned2018-03-29T01:08:31Z
dc.date.available2018-03-29T01:08:31Z
dc.identifierJornal De Pediatria. v. 85, n. 3, p. 217-22
dc.identifier1678-4782
dc.identifierdoi:10.2223/JPED.1886
dc.identifierhttp://www.ncbi.nlm.nih.gov/pubmed/19308296
dc.identifierhttp://repositorio.unicamp.br/jspui/handle/REPOSIP/198233
dc.identifier19308296
dc.identifier.urihttp://repositorioslatinoamericanos.uchile.cl/handle/2250/1298466
dc.descriptionTo evaluate how different ways of handling the neonatal self-inflating bag influences peak pressure and tidal volume. This is an experimental study involving 141 different professionals (physicians, resident physicians, physiotherapists, nurses and nursing technicians), who ventilated an artificial lung, adjusted to simulate the lung of a term neonate, using a self-inflating bag. Each professional handled the ventilator in five different ways: a) using both hands (10 fingers); and, with only one hand, b) five fingers, c) four fingers, d) three fingers and e) two fingers. Peak pressure and tidal volume data were recorded by the artificial lung equipment. Both variables showed high variability, from 2.5 to 106.3 cmH2O (mean = 39.73 cmH2O; 95%CI 37.32-42.13) for peak pressure, and from to 4 to 88 mL (mean = 39.56 mL; 95%CI 36.86-42.25) for tidal volume. There was no significant influence of the profession on any of the variables (p > 0.05). However, bag handling significantly influenced both peak pressure and tidal volume (p < 0.0001), which were higher when the operator used both hands. The results indicate that most professionals delivered excessively high peak pressures and tidal volumes, which could increase the risk of barotrauma and volutrauma, especially when both hands were used to ventilate. On the other hand, a small number of professionals delivered insufficient pressure and volume for adequate lung expansion and ventilation. The delivery of inadequate ventilation was not dependent on profession.
dc.description85
dc.description217-22
dc.languageeng
dc.languagepor
dc.relationJornal De Pediatria
dc.relationJ Pediatr (Rio J)
dc.rightsaberto
dc.rights
dc.sourcePubMed
dc.subjectArtificial Organs
dc.subjectHand
dc.subjectHumans
dc.subjectInfant, Newborn
dc.subjectInspiratory Capacity
dc.subjectLung
dc.subjectPressure
dc.subjectRespiration, Artificial
dc.subjectStatistics, Nonparametric
dc.subjectTerm Birth
dc.subjectTidal Volume
dc.subjectVentilators, Mechanical
dc.titlePeak Pressure And Tidal Volume Are Affected By How The Neonatal Self-inflating Bag Is Handled.
dc.typeArtículos de revistas


Este ítem pertenece a la siguiente institución