dc.creatorCosta E.T.
dc.creatorVieira Jr. F.U.
dc.creatorDe Medeiros Jr. J.D.
dc.creatorAntunes N.
dc.creatorVieira R.W.
dc.date2013
dc.date2015-06-25T19:09:38Z
dc.date2015-11-26T15:07:40Z
dc.date2015-06-25T19:09:38Z
dc.date2015-11-26T15:07:40Z
dc.date.accessioned2018-03-28T22:18:05Z
dc.date.available2018-03-28T22:18:05Z
dc.identifier9783642211973
dc.identifierIfmbe Proceedings. , v. 33 IFMBE, n. , p. 754 - 757, 2013.
dc.identifier16800737
dc.identifier10.1007/978-3-642-21198-0_192
dc.identifierhttp://www.scopus.com/inward/record.url?eid=2-s2.0-84875244832&partnerID=40&md5=aa2372b2b401936e3b560f8ffafc0760
dc.identifierhttp://www.repositorio.unicamp.br/handle/REPOSIP/88338
dc.identifierhttp://repositorio.unicamp.br/jspui/handle/REPOSIP/88338
dc.identifier2-s2.0-84875244832
dc.identifier.urihttp://repositorioslatinoamericanos.uchile.cl/handle/2250/1257514
dc.descriptionRoller pumps play an important role in extracorporeal blood circulation. However, pump occlusion must be properly performed and be adjusted mainly by the static and the dynamic methods. Our objective is to show preliminary results of the use of an auxiliary device for roller pump calibration via the dynamic method. Simultaneous tests have been carried out with the auxiliary device and a commercial data acquisition board and three disposable pressure transducers. The device has been also tested by perfusionists during the XXVIII Congresso Brasileiro de Circulação Extracorpórea. A roller pump and needed accessories for calibration via both the static and the dynamic methods have been made available to 56 perfusionists that were asked to calibrate the roller pump as they normally do in their hospital service. After each regulation the respective mean pressure of dynamic calibration was measured with our device. Results: A comparison between our device and the commercial data acquisition board has shown statistical equality for the adjusts of dynamic calibration pressure ( p > 0.05). The mean of the 56 measurements of dynamic calibration pressure was 434 ± 214 mmHg (mean ± standard deviation) and variation coefficient was VC = 49.2 %. Our results have shown that the constructed device may be used by perfusionists to quickly adjust roller pumps with great precision and repeatability. Careful tests may be initiated with our device in operating room. © 2013 Springer.
dc.description33 IFMBE
dc.description
dc.description754
dc.description757
dc.descriptionSKAN AG,Materialise NV,Bavarian Minist. Econ., Aff., Infrastruct., Transp. Technol.,Bavarian Bureau for International Business Relations GmbH,AKUD - Med Pharm GmbH
dc.descriptionGomes, W.J., Mendonca, J.T., Braile, D.M., Resultados em cirurgia cardiovascular. Oportunidade para rediscutir o atendimento médico e cardiológico no sistema público de saúde do país (2007) Rev Bras Cir Cardiovasc, 22 (4), pp. III-VI
dc.descriptionHaddad, N., Bittar, E., De Marchi, A.F., Hospital costs of coronary artery bypass grafting on elective coronary patients (2007) Arq Bras Cardiol, 88 (4), pp. 418-423
dc.descriptionMurray, C.J.L., Lopez, A.D., (1996) The Global Burden of Disease: A Comprehensive Assessment of Mortality and Disability from Disease, Injuries and Risk Factors in 1990 and Projected to 2020, , Boston: Harvard School of Public Health
dc.descriptionTayama, E., Teshima, H., Takaseya, T., Non-occlusive condition with the better-header roller pump: Impacts of flow dynamics and hemolysis (2004) Ann Thorac Cardiovasc Surg, 10 (6), pp. 357-361
dc.descriptionHansbro, S.D., Sharpe, D.A.C., Catchpole, R., Welsh, K.R., Munsch, C.M., McGoldrick, J.P., Kay, P.H., Haemolysis during cardiopulmonary bypass: An in vivo comparison of standard roller pumps, nonocclusive roller pumps and centrifugal pumps (1999) Perfusion, 14 (1), pp. 3-10
dc.descriptionAssant-Siaw, J., Tyrrel, J., Hotchinsky, A., Does the use of a centrifugal pump offer any additional benefit for patient having open heart surgery? (2006) Interact Cardio Vasc Thorac Surg, 5, pp. 128-134
dc.descriptionDenton, A., Cooley, M.D., Development of the roller pump for use in the cardiopulmonary bypass circuit (1987) Texas Heart Inst J, 14 (2), pp. 113-118
dc.descriptionTamari, Y., Lee-Sensiba, K., Leonard, E.F., Tortolani, A.J., A dynamic method for setting roller pumps nonocclusively reduces hemolysis and predicts retrograde flow (1997) ASAIO Journal, 43 (1), pp. 39-52. , DOI 10.1097/00002480-199701000-00009
dc.descriptionVieira Jr., F.U., Vieira, R.W., Antunes, N., The influence of the residual stress in silicone tubes in the calibration methods of roller pumps used in cardiopulmonary bypass (2010) ASAIO J, 56 (1), pp. 12-16
dc.descriptionVieira Jr., F.U., Vieira, R.W., Antunes, N., Analysis of hydrodynamic profile in different roller pumps models used in cardiopulmonary bypass (2009) Rev Bras Cir Cardiovasc, 24 (2), pp. 188-193
dc.descriptionVieira Jr., F.U., Vieira, R.W., Antunes, N., Considerações sobre métodos de ajuste de bombas de roletes usadas em circulação extracorpórea: Velocidade de queda e calibração dinâmica (2010) Rev Bras Eng Biom, 26 (1), pp. 25-32
dc.languagept
dc.publisher
dc.relationIFMBE Proceedings
dc.rightsfechado
dc.sourceScopus
dc.titleTestes Preliminares De Uso De Um Dispositivo Auxiliar Para O Ajuste De Bombas De Roletes Utilizadas Em Circulação Extracorpórea