dc.descriptionAlthough the self-inflating bag is widely used in the hospital setting, variability of delivered ventilatory parameters is usually high, which might result in both hypoventilation and lung injury. The aims of this study were to assess possible sources of the high variability and to evaluate the adequacy of obtained values in relation to the recommended values for neonatal resuscitation. METHODS: This was an experimental study in which 172 health professionals (physicians, resident physicians, physiotherapists, nurses, and nursing technicians) who work with neonatal intensive care manually ventilated a test lung (adjusted to simulate the lungs of an intubated term newborn) with a self-inflating bag in 5 different handling techniques, using 10, 5, 4, 3, and 2 fingers. Delivered values of peak inspiratory pressure (PIP), tidal volume (V T), and ventilatory frequency (f) were compared, taking into account the different handling modalities and professions by analysis of variance for repeated measures. Chi-square, the Friedman test and the Fisher exact tests were performed to compare the delivered and standard values. RESULTS: PIP and V T were significantly affected by the handling technique, with higher values for a greater number of fingers used for ventilation. Profession also influenced VT and f significantly: physiotherapists tended to deliver higher volumes and lower rates. Nevertheless, we observed high variability of all studied ventilatory parameters and overall inadequacy of obtained values. Most volunteers delivered excessive pressures and volumes at insufficient ventilatory frequency. CONCLUSIONS: Delivered values seem to depend on operators' individual and professional differences, as well as on the number of fingers used to compress the bag. However, from the clinical point of view, it is important to point out the high occurrence of inadequate delivered values, regardless of handling technique and profession. © 2012 Daedalus Enterprises.
dc.description57
dc.description4
dc.description525
dc.description530
dc.descriptionAlmeida, M.F., Guinsburg, R., Controversies in neonatal resuscitation (2001) J Pediatr (Rio J), 77 (SUPPL. 1), pp. S41-S52
dc.description2005 international consensus on cardiopulmonary rescue and emergency cardiovascular care sciences with treatment recommendations. Part 7: Neonatal resuscitation (2005) Resuscitation, 67 (2-3), pp. 293-303. , International Liaison Committee on Resuscitation
dc.descriptionGuinsburg, G., Almeida, M.F., Miyoshi, M.H., (2009) Manual de Reanimação Neonatal, pp. 125-238. , 5th edition, American Heart Association
dc.descriptionAmerican Academy of Pediatrics
dc.descriptionUniversidade Federal de São Paulo, editores, São Paulo: Editora Universidade Federal de São Paulo
dc.descriptionBook in Portugese
dc.descriptionHess, D., Goff, G., Johnson, K., The effect of hand size, resuscitator brand, and use of two hands on volumes delivered during adult bagvalve ventilation (1989) Respir Care, 34 (9), pp. 805-810
dc.descriptionHussey, S.G., Ryan, C.A., Murphy, B.P., Comparison of three manual ventilation devices using an intubated mannequin (2004) Arch Dis Child Fetal Neonatal Ed, 89 (3), pp. 490-493
dc.descriptionAugustine, J.A., Seidel, D.R., McCabe, J.B., Ventilation performance using a self-inflating anesthesia bag: Effect of operator characteristics (1987) Am J Emerg Med, 5 (4), pp. 267-270
dc.descriptionHess, D., Spahr, C., An evaluation of volumes delivered by selected adult disposable resuscitators: The effects of hand size, number of hands used and use of disposable medical gloves (1990) Respir Care, 35 (8), pp. 800-805
dc.descriptionHess, D., Goff, G., The effects of two-hand versus one-hand ventilation on volumes delivered during bag-valve ventilation at various resistances end compliances (1987) Respir Care, 32 (11), pp. 1025-1028
dc.descriptionHess, D., Simmons, M., Blaukovitch, S., Lightner, D., Doyle, T., An evaluation of the effects of fatigue, impedance, and use of two hands on volumes delivered during bag-valve ventilation (1993) Respir Care, 38 (3), pp. 271-275
dc.descriptionKain, Z.N., Berde, C.B., Benjamin, P.K., Thompson, J.E., Performance of pediatric resuscitation bags assessed with an infant lung simulator (1993) Anesth Analg, 77 (2), pp. 261-264
dc.descriptionBassani, M.A., Mezzacappa Filho, F., Coppo, M.R.C., Marba, S.T.M., Peak pressure and tidal volume are influenced by how the neonatal selfinflating bag is handled (2009) J Pediatr (Rio J), 85 (3), pp. 217-222
dc.descriptionLaw, G.D., Effect of hand size on VE, VT, and FiO2 during manual resuscitation (abstract) (1982) Respir Care, 27 (10), pp. 1236-1238
dc.descriptionFiner, N.N., Rich, W., Craft, A., Henderson, C., Comparison of methods of bag and mask ventilation for neonatal resuscitation (2001) Resuscitation, 49 (3), pp. 299-305
dc.descriptionDenehy, L., The use of manual hyperinflation in airway clearance (1999) Eur Respir J, 14 (4), pp. 958-965
dc.descriptionMcCarren, B., Chow, C.M., Manual hyperinflation: A description of the technique (1996) Aust J Physiother, 42 (3), pp. 203-208
dc.descriptionTurki, M., Young, M., Wagers, S.S., Bates, J.H.T., Peak pressures during manual ventilation (2005) Respir Care, 50 (3), pp. 340-344
dc.descriptionRezende, J.G., Menezes, C.G., Paula, A.M.C., Ferreira, A.C.P., Zaconeta, C.A.M., Silva, C.A.M., Avaliação do pico de pressão e de frequência respiratória durante o uso de balão auto-inflável em um modelo de pulmão neonatal (2006) J Pediatr (Rio J), 82 (5), pp. 359-364. , Article in Portugese
dc.descriptionRezende, J.G., Zaconeta, C.A.M., Ferreira, A.C.P., Silva, C.A.M., Rodrigues, M.P., Rebello, C.M., Tavares, P., Avaliação do pico de pressão, do volume corrente e da frequência respiratória durante ventilação de carneiros prematuros, utilizando balão auto-inflável (2006) J Pediatr (Rio J), 82 (4), pp. 279-283. , Article in Portugese
dc.description(2007) Standard specifications for performance and safety requirements for resuscitators intended for use with humans, , American Society for Testing and Materials, Designation: F920-93. Philadelphia: ASTM
dc.description1999. (Withdrawn)
dc.descriptionParadine, C.G., Rivett, B.H.P., (1974) Métodos estatísticos para tecnologias, p. 337. , São Paulo: Polígono da Editora da Universidade de São Paulo, Article in Portugese
dc.descriptionWyllie, J., Resuscitation of the term and premature baby (2008) Paediatr Child Health, 18 (4), pp. 166-171
dc.descriptionSpitzer, A.R., Fox, W.W., Positive-pressure ventilation: Pressure-limited and time-cycled ventilators (1996) Assisted ventilation of the neonate, pp. 167-186. , In: Goldsmith JP, Karotkin EH, editors, Philadelphia: Saunders
dc.descriptionGuinsburg, R., Almeida, M.F.B., Miyoshi, M.H., Reanimação do recémnascido em sala de parto (2004) Diagnóstico e tratamento em neonatologia, pp. 31-37. , In: Kopelman BI, Santos AMN, Goulart AL, Almeida MFB, Miyoshi MH, Guinsburg R, editores, São Paulo: Atheneu;:. Book in Portugese
dc.descriptionGanga-Zandzou, P.S., Diependaele, J.F., Riou, Y., Klosowski, S., Rakza, T., Logier, R., Lequien, P., La ventilation à Ambúchez nouveau-né: Une simple question de doigté? (1996) Arch Pédiatr, 3 (12), pp. 1270-1272. , Article in French
dc.descriptionRoehr, C.C., Kelm, M., Fischer, H.S., Bührer, C., Schmalisch, G., Proquitté, H., Manual ventilation devices in neonatal resuscitation: Tidal volume and positive pressure-provision (2010) Resuscitation, 81 (2), pp. 202-205
dc.descriptionGoldstein, B., Catlin, E., Vetere, J., Arguin, L., The role of in-line manometers in minimizing peak and mean airway pressure during the hand-regulated ventilation of newborn infants (1989) Respir Care, 34 (1), pp. 23-27
dc.descriptionFiner, N.N., Barrington, K.J., Al-Fadley, F., Peters, K.L., Limitations of self-inflating resuscitators (1986) Pediatrics, 77 (3), pp. 417-420
dc.descriptionGodoy, A.C., Vieira, R.J., de Capitani, E.M., Alterações de pico inspiratório e do volume corrente fornecidos por reanimadores manuais com balão auto-inflável em função do fluxo de entrada de oxigênio utilizado (2008) J Bras Pneumol, 34 (10), pp. 817-821. , Article in Portugese
dc.descriptionMcHale, S., Thomas, M., Hayden, E., Bergin, K., McCallion, N., Molloy, E.J., Variation in inspiratory time and tidal volume with T-piece neonatal resuscitator: Association with operator experience and distraction (2008) Resuscitation, 79 (2), pp. 230-233
dc.languageen
dc.publisher
dc.relationRespiratory Care
dc.rightsfechado
dc.sourceScopus
dc.titleAn Evaluation Of Peak Inspiratory Pressure, Tidal Volume, And Ventilatory Frequency During Ventilation With A Neonatal Self-inflating Bag Resuscitator
dc.typeArtículos de revistas


Este ítem pertenece a la siguiente institución