dc.creator | Bassani M.A. | |
dc.creator | Filho F.M. | |
dc.creator | Coppo M.R.C. | |
dc.creator | Marba S.T.M. | |
dc.date | 2012 | |
dc.date | 2015-06-26T20:29:23Z | |
dc.date | 2015-11-26T14:25:58Z | |
dc.date | 2015-06-26T20:29:23Z | |
dc.date | 2015-11-26T14:25:58Z | |
dc.date.accessioned | 2018-03-28T21:28:47Z | |
dc.date.available | 2018-03-28T21:28:47Z | |
dc.identifier | | |
dc.identifier | Respiratory Care. , v. 57, n. 4, p. 525 - 530, 2012. | |
dc.identifier | 201324 | |
dc.identifier | 10.4187/respcare.01423 | |
dc.identifier | http://www.scopus.com/inward/record.url?eid=2-s2.0-84860505664&partnerID=40&md5=36d5f8999f5ab2281bdc76993c328f49 | |
dc.identifier | http://www.repositorio.unicamp.br/handle/REPOSIP/97012 | |
dc.identifier | http://repositorio.unicamp.br/jspui/handle/REPOSIP/97012 | |
dc.identifier | 2-s2.0-84860505664 | |
dc.identifier.uri | http://repositorioslatinoamericanos.uchile.cl/handle/2250/1245921 | |
dc.description | Although the self-inflating bag is widely used in the hospital setting, variability of delivered ventilatory parameters is usually high, which might result in both hypoventilation and lung injury. The aims of this study were to assess possible sources of the high variability and to evaluate the adequacy of obtained values in relation to the recommended values for neonatal resuscitation. METHODS: This was an experimental study in which 172 health professionals (physicians, resident physicians, physiotherapists, nurses, and nursing technicians) who work with neonatal intensive care manually ventilated a test lung (adjusted to simulate the lungs of an intubated term newborn) with a self-inflating bag in 5 different handling techniques, using 10, 5, 4, 3, and 2 fingers. Delivered values of peak inspiratory pressure (PIP), tidal volume (V T), and ventilatory frequency (f) were compared, taking into account the different handling modalities and professions by analysis of variance for repeated measures. Chi-square, the Friedman test and the Fisher exact tests were performed to compare the delivered and standard values. RESULTS: PIP and V T were significantly affected by the handling technique, with higher values for a greater number of fingers used for ventilation. Profession also influenced VT and f significantly: physiotherapists tended to deliver higher volumes and lower rates. Nevertheless, we observed high variability of all studied ventilatory parameters and overall inadequacy of obtained values. Most volunteers delivered excessive pressures and volumes at insufficient ventilatory frequency. CONCLUSIONS: Delivered values seem to depend on operators' individual and professional differences, as well as on the number of fingers used to compress the bag. However, from the clinical point of view, it is important to point out the high occurrence of inadequate delivered values, regardless of handling technique and profession. © 2012 Daedalus Enterprises. | |
dc.description | 57 | |
dc.description | 4 | |
dc.description | 525 | |
dc.description | 530 | |
dc.description | Almeida, M.F., Guinsburg, R., Controversies in neonatal resuscitation (2001) J Pediatr (Rio J), 77 (SUPPL. 1), pp. S41-S52 | |
dc.description | 2005 international consensus on cardiopulmonary rescue and emergency cardiovascular care sciences with treatment recommendations. Part 7: Neonatal resuscitation (2005) Resuscitation, 67 (2-3), pp. 293-303. , International Liaison Committee on Resuscitation | |
dc.description | Guinsburg, G., Almeida, M.F., Miyoshi, M.H., (2009) Manual de Reanimação Neonatal, pp. 125-238. , 5th edition, American Heart Association | |
dc.description | American Academy of Pediatrics | |
dc.description | Universidade Federal de São Paulo, editores, São Paulo: Editora Universidade Federal de São Paulo | |
dc.description | Book in Portugese | |
dc.description | Hess, D., Goff, G., Johnson, K., The effect of hand size, resuscitator brand, and use of two hands on volumes delivered during adult bagvalve ventilation (1989) Respir Care, 34 (9), pp. 805-810 | |
dc.description | Hussey, S.G., Ryan, C.A., Murphy, B.P., Comparison of three manual ventilation devices using an intubated mannequin (2004) Arch Dis Child Fetal Neonatal Ed, 89 (3), pp. 490-493 | |
dc.description | Augustine, J.A., Seidel, D.R., McCabe, J.B., Ventilation performance using a self-inflating anesthesia bag: Effect of operator characteristics (1987) Am J Emerg Med, 5 (4), pp. 267-270 | |
dc.description | Hess, D., Spahr, C., An evaluation of volumes delivered by selected adult disposable resuscitators: The effects of hand size, number of hands used and use of disposable medical gloves (1990) Respir Care, 35 (8), pp. 800-805 | |
dc.description | Hess, D., Goff, G., The effects of two-hand versus one-hand ventilation on volumes delivered during bag-valve ventilation at various resistances end compliances (1987) Respir Care, 32 (11), pp. 1025-1028 | |
dc.description | Hess, D., Simmons, M., Blaukovitch, S., Lightner, D., Doyle, T., An evaluation of the effects of fatigue, impedance, and use of two hands on volumes delivered during bag-valve ventilation (1993) Respir Care, 38 (3), pp. 271-275 | |
dc.description | Kain, Z.N., Berde, C.B., Benjamin, P.K., Thompson, J.E., Performance of pediatric resuscitation bags assessed with an infant lung simulator (1993) Anesth Analg, 77 (2), pp. 261-264 | |
dc.description | Bassani, M.A., Mezzacappa Filho, F., Coppo, M.R.C., Marba, S.T.M., Peak pressure and tidal volume are influenced by how the neonatal selfinflating bag is handled (2009) J Pediatr (Rio J), 85 (3), pp. 217-222 | |
dc.description | Law, G.D., Effect of hand size on VE, VT, and FiO2 during manual resuscitation (abstract) (1982) Respir Care, 27 (10), pp. 1236-1238 | |
dc.description | Finer, N.N., Rich, W., Craft, A., Henderson, C., Comparison of methods of bag and mask ventilation for neonatal resuscitation (2001) Resuscitation, 49 (3), pp. 299-305 | |
dc.description | Denehy, L., The use of manual hyperinflation in airway clearance (1999) Eur Respir J, 14 (4), pp. 958-965 | |
dc.description | McCarren, B., Chow, C.M., Manual hyperinflation: A description of the technique (1996) Aust J Physiother, 42 (3), pp. 203-208 | |
dc.description | Turki, M., Young, M., Wagers, S.S., Bates, J.H.T., Peak pressures during manual ventilation (2005) Respir Care, 50 (3), pp. 340-344 | |
dc.description | Rezende, J.G., Menezes, C.G., Paula, A.M.C., Ferreira, A.C.P., Zaconeta, C.A.M., Silva, C.A.M., Avaliação do pico de pressão e de frequência respiratória durante o uso de balão auto-inflável em um modelo de pulmão neonatal (2006) J Pediatr (Rio J), 82 (5), pp. 359-364. , Article in Portugese | |
dc.description | Rezende, J.G., Zaconeta, C.A.M., Ferreira, A.C.P., Silva, C.A.M., Rodrigues, M.P., Rebello, C.M., Tavares, P., Avaliação do pico de pressão, do volume corrente e da frequência respiratória durante ventilação de carneiros prematuros, utilizando balão auto-inflável (2006) J Pediatr (Rio J), 82 (4), pp. 279-283. , Article in Portugese | |
dc.description | (2007) Standard specifications for performance and safety requirements for resuscitators intended for use with humans, , American Society for Testing and Materials, Designation: F920-93. Philadelphia: ASTM | |
dc.description | 1999. (Withdrawn) | |
dc.description | Paradine, C.G., Rivett, B.H.P., (1974) Métodos estatísticos para tecnologias, p. 337. , São Paulo: Polígono da Editora da Universidade de São Paulo, Article in Portugese | |
dc.description | Wyllie, J., Resuscitation of the term and premature baby (2008) Paediatr Child Health, 18 (4), pp. 166-171 | |
dc.description | Spitzer, A.R., Fox, W.W., Positive-pressure ventilation: Pressure-limited and time-cycled ventilators (1996) Assisted ventilation of the neonate, pp. 167-186. , In: Goldsmith JP, Karotkin EH, editors, Philadelphia: Saunders | |
dc.description | Guinsburg, R., Almeida, M.F.B., Miyoshi, M.H., Reanimação do recémnascido em sala de parto (2004) Diagnóstico e tratamento em neonatologia, pp. 31-37. , In: Kopelman BI, Santos AMN, Goulart AL, Almeida MFB, Miyoshi MH, Guinsburg R, editores, São Paulo: Atheneu;:. Book in Portugese | |
dc.description | Ganga-Zandzou, P.S., Diependaele, J.F., Riou, Y., Klosowski, S., Rakza, T., Logier, R., Lequien, P., La ventilation à Ambúchez nouveau-né: Une simple question de doigté? (1996) Arch Pédiatr, 3 (12), pp. 1270-1272. , Article in French | |
dc.description | Roehr, C.C., Kelm, M., Fischer, H.S., Bührer, C., Schmalisch, G., Proquitté, H., Manual ventilation devices in neonatal resuscitation: Tidal volume and positive pressure-provision (2010) Resuscitation, 81 (2), pp. 202-205 | |
dc.description | Goldstein, B., Catlin, E., Vetere, J., Arguin, L., The role of in-line manometers in minimizing peak and mean airway pressure during the hand-regulated ventilation of newborn infants (1989) Respir Care, 34 (1), pp. 23-27 | |
dc.description | Finer, N.N., Barrington, K.J., Al-Fadley, F., Peters, K.L., Limitations of self-inflating resuscitators (1986) Pediatrics, 77 (3), pp. 417-420 | |
dc.description | Godoy, A.C., Vieira, R.J., de Capitani, E.M., Alterações de pico inspiratório e do volume corrente fornecidos por reanimadores manuais com balão auto-inflável em função do fluxo de entrada de oxigênio utilizado (2008) J Bras Pneumol, 34 (10), pp. 817-821. , Article in Portugese | |
dc.description | McHale, S., Thomas, M., Hayden, E., Bergin, K., McCallion, N., Molloy, E.J., Variation in inspiratory time and tidal volume with T-piece neonatal resuscitator: Association with operator experience and distraction (2008) Resuscitation, 79 (2), pp. 230-233 | |
dc.language | en | |
dc.publisher | | |
dc.relation | Respiratory Care | |
dc.rights | fechado | |
dc.source | Scopus | |
dc.title | An Evaluation Of Peak Inspiratory Pressure, Tidal Volume, And Ventilatory Frequency During Ventilation With A Neonatal Self-inflating Bag Resuscitator | |
dc.type | Artículos de revistas | |