dc.identifier | Pino Garcés, Catherin Alicia. (2017). Comisionamiento y puesta en marcha del equipo de braquiterapia de alta tasa de dosis (microselectron) en la Unidad de Radioterapia del Hospital Carlos Andrade Marín, para tratamiento de pacientes con cáncer Ginecológico. Escuela Superior Politécnica de Chimborazo. Riobamba. | |
dc.identifier | http://dspace.espoch.edu.ec/handle/123456789/6913 | |
dc.identifier.uri | http://repositorioslatinoamericanos.uchile.cl/handle/2250/1063624 | |
dc.description | El objetivo fue comisionar y poner en marcha el equipo de braquiterapia de alta tasa de dosis
(microSelectron) de la Unidad de Radioterapia del Hospital Carlos Andrade Marín, para el
tratamiento de pacientes con cáncer ginecológico. Se realizó un análisis del reporte Task Group
No.43 de la Asociación Americana de Físicos Médicos (AAPM), reporte TECDOC 1151 basado
en los Aspectos físicos de garantía de calidad en radioterapia: Protocolo de control de calidad del
Organismo Internacional de Energía Atómica (OIEA), y los manuales de usuario del equipo
microSelectron-HDR. Se utilizó el microSelectron-HDR, la cámara de pozo, detectores de
radiación, electrómetro, guías de conducción para las mediciones, mediante el TECDOC 1151 y
el manual de Braquiterapia de Alta tasa de dosis (BATD) del Instituto Nacional del Cáncer
(INCA) – Brasil, teniendo en cuenta el posicionamiento de los instrumentos de medición y varios
factores como sensibilidad, radiación de fuga, reproducibilidad y repetividad, entre otros.
Posteriormente se calibraron los equipos del conjunto dosimétrico mediante la utilización del
Acelerador lineal Sinergy con una energía de 6MV y 100UM; continuando con la determinación
de las pruebas necesarias para el control de calidad del equipo de alta tasa de dosis. El error
obtenido una vez calibrada la fuente de Iridio 192 por medio de la cámara de pozo fue del 3%,
tolerancia de ±1 con respecto al posicionamiento de la fuente con filme radiográfico y tiempo de
permanencia de la fuente una tolerancia de ±0.5 con respecto a las cartas de control que indican
la estabilidad del proceso. Se concluye que el microSelectron-HDR se encuentra listo para uso
clínico en el tratamiento de pacientes con cáncer ginecológico, debido a que cumple con todos los
parámetros recomendados por los protocolos internacionales, se recomienda utilizar los
resultados de este trabajo como referencia si se requiere realizar un control de calidad del equipo
para verificar su funcionalidad y límites de tolerancia. | |
dc.description | The objective was to commission and implement the high dose rate of brachytherapy
equipment (microSelectron) of Carlos Andrade Marín Hospital Radiotherapy Unit for the
treatment of patients with gynecological cancer. A report was made by Task Group N.43
of the American Association of Physicists in Medicine (AAPM), TECDOC 1151 report
based on physical aspects of quality assurance in radiotherapy: Quality control protocol
of the International Atomic Energy Agency (IAEA), and the microSelectron-HDR user
manuals.
The microSelectron-HDR, the well chamber, radiation detectors, electrometer,
conduction guides for the measurements were used, through the TECDOC 1151 and the
Manual of High Rate of Brachytherapy (BATD) of de National Cancer Institute (INCA)
–Brazil. Taking into account the positioning of measuring instruments and various factors
such as sensitivity, leakage radiation, reproducibility and repeatability, among others.
Later, the dosimetric set equipment was calibrated by using the Linear Accelerator
Synergy with an energy of 6MV and 100 UM; continuing with the determination of
necessary tests for the equipment quality control of high dose rate. The error obtained
after calibrating the source of Iridium 192 by means of the well chamber was 3% tolerance
of ±1 with respect to the positioning of the source with radiographic film and time of
permanence of the source a tolerance of ±0.5 with respect to control charts indicating the
process stability. It is concluded that the microSelectron-HDR is ready for clinical use in
the treatment of patients with gynecological cancer, because it complies with all the
parameters recommended by the international protocols. It is recommended to use the
results of this work as reference if it is necessary to perform a quality control of the
equipment to verify its functionality and limits of tolerance. | |
dc.language | spa | |
dc.publisher | Escuela Superior Politécnica de Chimborazo | |
dc.relation | UDCTFC;86T00040 | |
dc.rights | openAccess | |
dc.rights | http://creativecommons.org/licenses/by-nc-sa/3.0/ec/ | |
dc.subject | TECNOLOGÍA Y CIENCIAS MÉDICAS | |
dc.subject | BIOFÍSICA | |
dc.subject | RADIOTERAPIA | |
dc.subject | MICROSELECTRON-HDR | |
dc.subject | BRAQUITERAPIA INTRACAVITARIA | |
dc.subject | ONCENTRA (SOFTWARE) | |
dc.subject | IRIDIO-192 | |
dc.subject | CONTROL DE CALIDAD | |
dc.title | Comisionamiento y puesta en marcha del equipo de braquiterapia de alta tasa de dosis (microselectron) en la Unidad de Radioterapia del Hospital Carlos Andrade Marín, para tratamiento de pacientes con cáncer Ginecológico. | |
dc.type | Tesis | |