Artigo de peri??dico
Variability
Registro en:
0969-8043
160
10.1016/j.apradiso.2020.109108
0000-0002-4049-6720
0000-0003-4023-236X
0000-0003-2281-5756
0000-0001-9475-8266
34.96
44.00
Autor
SILVA J??NIOR, IREMAR A. da
SIQUEIRA, PAULO de T.D.
NASCIMENTO, EDUARDO do
YORIYAZ, HELIO
SORDI, GIAN-MARIA A.A.
POTIENS, MARIA da P.A.
Resumen
In this paper we present the impact of variability, a surface source parameter, on the efficiency evaluation of
surface contamination monitors. This study was based on two source uniformity correction methodologies and
data from real surface source distributions. Surface source intensity distribution has been changed by rearranging
the cells (portions of the active area of each LARS) while keeping the same source uniformity value. Instrument
efficiencies have been calculated for different sets of uniformities and variabilities. This study led to emphasize
the importance of variability, a differential source intensity distribution parameter, over the uniformity, an integral
source intensity distribution parameter, and reinforced the importance of the source uniformity correction
procedure on the course of surface contamination monitor calibration. Funda????o de Amparo ?? Pesquisa do Estado de S??o Paulo (FAPESP) FAPESP: 08/57863-2