Conductometric titration of Ag+ with thioacetamide
Titulação condutométrica de Ag+ com tioacetamina
Registro en:
10.5902/2179460X24916
Autor
Schneider, Ilzemaro
Pedroso, Antonio Gilberto
Silva, Paulo Augusto M. da
Institución
Resumen
This work shows the viability of conductometric titration of Ag+ in aqueous medium, with thioacetamide. First, the stability of thioacetamide in aqueous solution was tested. Analysis of during 5 weeks weeks did not show any concentration change in a thioacetamide solution.To get betteraccuracy in the results o four analysis, the titrating and titrated solutions were used in different concentrations. The lowest relative error (0.12%) was found in the titration of a 0.01N solution of silver nitrate with a 0.1 N solution of thioacetamide. Demonstrou-se a viabilidade da titulação condutométrica de Ag+ em meio aquoso, com a tioacetamida. O erro relativo foi de 0,12%.Inicialmente se comprovou a estabilidade da tioacetamida em solução aquosa. Análises efetuadas durante 5 semanas, a intervalos de uma semana, não acusaram alteração de concentração numa solução de tioacetamida.Para obter-se melhor precisão nas determinações, foram usa das diferentes concentrações de titulante e titulado. O menor erro relativo (0,12%) ocorreu com solução N de tioacetamida e solução 0,01 N de nitrato de prata.