info:eu-repo/semantics/article
Optimization of a UV-Visible Spectrophotometer Response Using a Fractional Factorial Design
Optimización de la respuesta de un espectrofotómetro UV-Visible utilizando un diseño factorial fraccional
Registro en:
10.18272/aci.v3i2.62
Autor
Cazar, Robert A.
Institución
Resumen
Three operation parameters of a UV-visible spectrophotometer (scan speed, monochromator slit width, and sampling interval) were optimized utilizing a fractional factorial design. Four absorption spectra of holmium oxide were recorded in a portion of the visible region (560-610 nanometers) using different combinations of "high" and "low" levels for the parameters already mentioned. The combinations were provided by a 23-1 design. From the results it is deduced that the highest quality spectrum is furnished by the following combination: high scan speed, narrow monochromator slit width (0.2 nm.) and small sampling interval (one data point per each 0.1 nanometer). This work illustrates the usefulness of fractional factorial design as an optimization tool in the analytical chemistry laboratory. Tres parámetros de operación de un espectrofotómetro UV-visible (velocidad de barrido, ancho de ranura del monocromador e intervalo de muestreo) fueron optimizados utilizando un diseño factorial fraccionado. Cuatro espectros de absorción de óxido de holmio fueron registrados en una porción de la región visible (560-610 nanómetros) usando diferentes combinaciones de niveles "altos" y "bajos" para los parámetros ya mencionados. Las combinaciones fueron provistas por un diseño 23-1. De los resultados se deduce que el espectro de más alta calidad es proporcionado por la siguiente combinación: alta velocidad de barrido, ancho de ranura del monocromador pequeño (0.2 nanómetros) e intervalo de muestreo pequeño (un punto por cada 0.1 nanómetro). Este trabajo ilustra la utilidad del diseño factorial fraccionado como una herramienta de optimización en el laboratorio de química analítica