dc.creatorNovoa, Carlos
dc.creatorKoga, Roberto
dc.creatorPortilla, Arturo
dc.creatorOtero, Manuel
dc.creatorMoore, Mariel
dc.creatorOcaña, Elías
dc.creatorParagulla, Wilson
dc.creatorHerrera, Jorge
dc.creatorMartínez, Ramos
dc.creatorRobles, Anita
dc.creatorMiranda, Jesús
dc.creatorCavero, Luis
dc.creatorCavero, Luis
dc.creatorRobles, Anita
dc.creatorMiranda, Jesús
dc.creatorMartínez, Ramos
dc.creatorParagulla, Wilson
dc.creatorMoore, Mariel
dc.creatorHerrera, Jorge
dc.creatorOcaña, Elías
dc.creatorPortilla, Arturo
dc.creatorOtero, Manuel
dc.creatorNovoa, Carlos
dc.creatorKoga, Roberto
dc.date.accessioned2016-11-23T20:00:38Z
dc.date.accessioned2023-05-24T15:13:53Z
dc.date.available2016-11-23T20:00:38Z
dc.date.available2023-05-24T15:13:53Z
dc.date.created2016-11-23T20:00:38Z
dc.date.issued2016-05
dc.identifierCavero L, Robles A, Miranda J, Martínez R, Paragulla W, Moore M, et al. Calificación de un recinto de producción y envasado de pertecnetato de sodio Tc 99m. Informe Científico Tecnológico. 2014; 14:201-208.
dc.identifier1684-1662
dc.identifierhttps://hdl.handle.net/20.500.13054/620
dc.identifier.urihttps://repositorioslatinoamericanos.uchile.cl/handle/2250/6442814
dc.description.abstractSe diseñó y aplicó un protocolo de calificación para un recinto de producción y envasado de pertecnetato de sodio Tc99m, en un proceso de dos etapas: calificación de la instalación (CI) y calificación de operación (CO). En la CI se verificó las especificaciones de diseño y los requerimientos del usuario, así como el equipamiento y materiales asociados mediante documentación trazable. En la CO se verificó las secuencias programadas para su operación y control, además se realizaron las pruebas operacionales recomendadas por la Organización Mundial de la Salud (OMS) según la norma ISO 14644-1 y 3 para áreas limpias. Los resultados obtenidos evidenciaron que el recinto cumple con la clasificación para los Grados C y A para la preparación y envasado de Tc 99m.
dc.description.abstractIt was designed and implemented a protocol for a hot cell of production and packaging of sodium pertechnetate Tc99m, in a two-step process: installation qualification (IQ) and operation qualification (OQ). In the IQ design specifications and user requirements and associated equipment and materials by traceable documentation was verified. In the OQ scheduled for operation and control sequences it was verified plus operational tests recommended by the World Health Organization (WHO) and ISO 14644-1 and 3 were performed to clean areas. The results showed that the hot cell complies with the classification for Grades C and A for the preparation and packaging of Tc 99m.
dc.languagespa
dc.publisherInstituto Peruano de Energía Nuclear
dc.publisherPE
dc.rightsinfo:eu-repo/semantics/openAccess
dc.sourceInstituto Peruano de Energía Nuclear
dc.sourceRepositorio Institucional del Instituto Peruano de Energía Nuclear
dc.subjectProducción de isótopos
dc.subjectTecnecio 99
dc.subjectPertecnetatos
dc.subjectCeldas calientes
dc.subjectNormalización
dc.subjectCompuestos de sodio
dc.subjectEmbalaje
dc.titleCalificación de un recinto de producción y envasado de pertecnetato de sodio Tc 99m
dc.typeinfo:eu-repo/semantics/article


Este ítem pertenece a la siguiente institución