dc.creatorWuilloud, Gustavo M.
dc.creatorWuilloud, Rodolfo German
dc.creatorAltamirano, Jorgelina Cecilia
dc.creatorOlsina, Roberto Antonio
dc.creatorMartinez, Luis Dante
dc.date.accessioned2020-03-11T20:34:21Z
dc.date.accessioned2022-10-15T11:55:50Z
dc.date.available2020-03-11T20:34:21Z
dc.date.available2022-10-15T11:55:50Z
dc.date.created2020-03-11T20:34:21Z
dc.date.issued2003-02
dc.identifierWuilloud, Gustavo M.; Wuilloud, Rodolfo German; Altamirano, Jorgelina Cecilia; Olsina, Roberto Antonio; Martinez, Luis Dante; On-line preconcentration and determination of chromium in parenteral solutions by flow injection - Flame atomic absorption spectrometry; Elsevier Science; Journal of Pharmaceutical and Biomedical Analysis; 31; 1; 2-2003; 117-124
dc.identifier0731-7085
dc.identifierhttp://hdl.handle.net/11336/99214
dc.identifierCONICET Digital
dc.identifierCONICET
dc.identifier.urihttps://repositorioslatinoamericanos.uchile.cl/handle/2250/4383278
dc.description.abstractAn on-line chromium preconcentration and determination system implemented with flame atomic absorption spectrometry (FAAS) associated to flow injection (FI) was studied. For the retention of chromium, 4-(2-Thiazolylazo)-resorcinol (TAR) and Amberlite XAD-16 were used, at pH 5.0. The Cr-TAR complex was removed from the micro-column with ethanol. An enrichment factor of 50 was obtained for the preconcentration of 50 ml of sample solution. The detection limit value for the preconcentration of 50 ml of aqueous solution of Cr was 20 ng l-1. The precision for ten replicate determinations at the 5 μg l-1 Cr levels was 2.9% relative standard deviation (RSD), calculated from the peak heights obtained. The calibration graph using the preconcentration system for chromium was linear with a correlation coefficient of 0.9997 at levels near the detection limits up to at least 100 μg l-1. The method was successfully applied to the determination of chromium in parenteral solution samples. © 2002 Elsevier Science B.V. All rights reserved.
dc.languageeng
dc.publisherElsevier Science
dc.relationinfo:eu-repo/semantics/altIdentifier/doi/http://dx.doi.org/10.1016/S0731-7085(02)00604-0
dc.relationinfo:eu-repo/semantics/altIdentifier/url/https://www.sciencedirect.com/science/article/pii/S0731708502006040
dc.rightshttps://creativecommons.org/licenses/by-nc-nd/2.5/ar/
dc.rightsinfo:eu-repo/semantics/restrictedAccess
dc.subject4-(2-THIAZOLYLAZO)-RESORCINOL
dc.subjectCHROMIUM IN PARENTERAL SOLUTIONS SAMPLES
dc.subjectFAAS
dc.subjectFLOW INJECTION
dc.subjectON-LINE PRECONCENTRATION
dc.titleOn-line preconcentration and determination of chromium in parenteral solutions by flow injection - Flame atomic absorption spectrometry
dc.typeinfo:eu-repo/semantics/article