info:eu-repo/semantics/article
Variation of the noise levels in the baseline of an adsorption microcalorimeter
Fecha
2009-08Registro en:
García Cuello, Vanessa Silenia; Moreno Piraján, Juan Carlos; Giraldo Gutiérrez, L.; Sapag, Manuel Karim; Zgrablich, Jorge Andres; Variation of the noise levels in the baseline of an adsorption microcalorimeter; Springer; Journal of Thermal Analysis and Calorimetry; 97; 2; 8-2009; 705-709
1388-6150
1588-2926
CONICET Digital
CONICET
Autor
García Cuello, Vanessa Silenia
Moreno Piraján, Juan Carlos
Giraldo Gutiérrez, L.
Sapag, Manuel Karim
Zgrablich, Jorge Andres
Resumen
This work shows the results obtained to determine the noise in the baseline of a specially designed Tian Calvet-Type adsorption microcalorimeter. The results show that noise levels vary from 0.5 to 10 μV, which were evaluated varying the electrical work and the micro calorimeter surrounding temperature. Relationships can be seen between the variables employed in the observation of stability, temperature, potency levels and generated noise. © 2009 Akadémiai Kiadó, Budapest, Hungary.