Skin-electrode impedance measurement during ECG acquisition: Method's validation
Fecha
2016-04Registro en:
Casal, Leonardo; la Mura, Guillermo Mario; Skin-electrode impedance measurement during ECG acquisition: Method's validation; IOP Publishing; Journal of Physics: Conference Series; 705; 1; 4-2016; 1-7
1742-6596
CONICET Digital
CONICET
Autor
Casal, Leonardo
la Mura, Guillermo Mario
Resumen
Skm-electrode impedance measurement can provide valuable information prior. dunng and post electrocardiographic (ECG) or electroencephalographs (EEG) acquisitions. In this work we validate a method for skm-electrode impedance measurement using test circuits with known resistance and capacitor values, at different frequencies for injected excitation current. Finally the method is successfully used for impedance measurement during ECG acquisition on a subject usmg 125 Hz and 6 nA square wave excitation signal at instrumentation amplifier mput. The method can be used for many electrodes configuration.