dc.creator | Ramírez E,Pablo | |
dc.creator | Barrero V,Raúl | |
dc.creator | González F,Rodrigo | |
dc.creator | Frangini S,Patricia | |
dc.creator | Vergara S,Ismael | |
dc.date | 2011-07-01 | |
dc.date.accessioned | 2017-03-07T16:39:01Z | |
dc.date.available | 2017-03-07T16:39:01Z | |
dc.identifier | http://www.scielo.cl/scielo.php?script=sci_arttext&pid=S0034-98872011000700014 | |
dc.identifier.uri | http://repositorioslatinoamericanos.uchile.cl/handle/2250/405213 | |
dc.description | We report a 26year old patient who had a single chamber pacemaker implantation one year before. During a routine pre-operative evaluation, pacemaker dysfunction was demonstrated due to sensing and pacing failure, associated to left pectoral muscle rhythmic contraction. Chest X-ray confirmed Twiddler syndrome, in which twisting or rotation of the device inside the pocket results in lead dislodgement and device malfunction. | |
dc.format | text/html | |
dc.language | es | |
dc.publisher | Sociedad Médica de Santiago | |
dc.source | Revista médica de Chile v.139 n.7 2011 | |
dc.subject | Arrhythmias, cardiac | |
dc.subject | Pacemaker, artificial | |
dc.subject | Twiddler syndrome | |
dc.title | Síndrome de Twiddler: una causa infrecuente de disfunción de marcapaso definitivo. Caso clínico | |
dc.type | Artículos de revistas | |