Otro
Implementación de un método de verificación de tratamientos IMRT utilizando archivos MLCLogs
Autor
Nieto Lenes, Javier Alberto
Institución
Resumen
In this work we investigated and implemented a method of verifying the dose distribution for intensity modulated radiotherapy (IMRT) using the DynaLogs files. DynaLogs files record information every 20 to 50ms of the planned and executed positions of the mechanical components of the linear accelerator. This information is used to reconstruct the administered fluence and recalculate the absorbed dose on the patient's anatomy. For the treatment plans selected in the present work, a maximum variation of 0.650% was found, which does not imply a significant variation between the planned dose and the dose delivered during treatment. The implementation of this method allows to validate IMRT treatments on the patient's anatomy. En este trabajo se investigó e implemento un método de verificación de la distribución de dosis para la radioterapia de intensidad modulada (IMRT) utilizando los archivos DynaLogs. Los archivos DynaLogs registran información cada 20 o 50ms de la posiciones planeadas y ejecutadas de los componentes mecánicos del acelerador lineal. Esta información es utilizada para reconstruir la fluencia administrada y recalcular la dosis absorbida sobre la anatomía del paciente. Para los planes de tratamiento seleccionados en el presente trabajo, se encontró una variación de 0.650% máximo, lo que no implica una variación significativa entre la dosis planeada y la dosis suministrada durante el tratamiento. La implementación de este método permite validar los tratamientos IMRT sobre la anatomía del paciente.