Artigo
Estudo mecânico de implante para fixação do segmento lombossacro da coluna vertebral
Fecha
2000-09-01Registro en:
Acta Ortopédica Brasileira. Sociedade Brasileira de Ortopedia e Traumatologia, v. 8, n. 3, p. 112-121, 2000.
1413-7852
S1413-78522000000300003.pdf
S1413-78522000000300003
10.1590/S1413-78522000000300003
Autor
Serdeira, Afrane [UNIFESP]
Barros Fº, Tarcísio Eloy Pessoa de
Puertas, Eduardo Barros [UNIFESP]
Laredo Filho, José [UNIFESP]
Leivas, Tomaz Puga
Institución
Resumen
This work consists of a mechanical analysis of the resistance and critical points of a stainless steel device for lumbossacral spinal fixation. The device was fixed into a wooden model representing the spinal lumbossacral segment. The experiment comprised seven tests of flexo-compression, seven tests of axial rigidity, seven tests of radial rigidity, and one destructive test. The critical points are the intersection of the pedicular screws threads and the attrition rate between the vise and the rod. The results demonstrated the device to be efficient and safe when used in human beings. Foram estudados, do ponto de vista mecânico, a rigidez e os pontos críticos de um implante para fixação interna da coluna lombossacra. Aplicou-se o implante sobre um modelo de madeira simulando o segmento lombossacro da coluna. Realizamos sete ensaios de flexo-compressão, sete de rigidez axial, sete de rigidez radial e um ensaio destrutivo. Os resultados demonstraram que o implante foi eficiente e seguro para uso em seres humanos.