Artículos de revistas
Fracture of Zirconia Abutment with Metallic Insertion on Anterior Single Titanium Implant with Internal Hexagon: Retrieval Analysis of a Failure
Fecha
2016-09-01Registro en:
European Journal Of Prosthodontics And Restorative Dentistry. Lowestoft: Dennis Barber Ltd, v. 24, n. 3, p. 164-168, 2016.
0965-7452
10.1922/EJPRD_01556Bottino05
WOS:000411710300007
9234456003563666
Autor
Universidade Estadual Paulista (Unesp)
Univ Zurich
Institución
Resumen
This case report presents a retrieval analysis of a screw-retained one-piece restoration with the veneering ceramic fired directly onto the zirconia abutment that fractured during insertion. A patient who experienced root fracture of a maxillary left central incisor received a titanium implant on the same day as extraction. After delayed implant loading, a two-piece zirconia abutment with metallic insertion was customized. Upon installation, a horizontal fracture of the crown just above the metallic portion was detected. The retrieval analysis of a fractured zirconia abutment showed crack formation and diffusion of glaze material that expanded the crack line during firing.