dc.contributorUniversidade Estadual Paulista (Unesp)
dc.contributorFaculdade de Americana
dc.date.accessioned2016-01-28T16:57:10Z
dc.date.available2016-01-28T16:57:10Z
dc.date.created2016-01-28T16:57:10Z
dc.date.issued2012
dc.identifierAmerican Journal of Analytical Chemistry, v. 3, n. 2, p. 113-117, 2012.
dc.identifier2156-8251
dc.identifierhttp://hdl.handle.net/11449/133927
dc.identifier10.4236/ajac.2012.32016
dc.identifier1427125996716282
dc.identifier3175417937754649
dc.identifier9881720291571774
dc.identifier1427125996716282
dc.description.abstractAn ion-exchange chromatography (IEC) followed by post-column derivatization method was validated for the determination of glutamine in food additive products. The chromatographic method was carried out on a sodium cation-exchange column and using detection at 570 nm. The mobile phase consisted of sodium eluant, pH 3.15, with 5% sulfolane, sodium eluant, pH 7.40 and sodium column regenerant. The method was linear in the range of 4 - 50 µg/mL (r 2 = 0.9999). The accuracy was 99.78%. Moreover, method validation demonstrated acceptable results for precision and robustness. The method was found to be able to use for routine analysis of the glutamine.
dc.languageeng
dc.publisherScientific Research
dc.relationAmerican Journal of Analytical Chemistry
dc.rightsAcesso restrito
dc.sourceCurrículo Lattes
dc.subjectGlutamine
dc.subjectFood additive product
dc.subjectLiquid chromatography
dc.titleValidation of a HPLC method for determination of glutamine in food additives using post-column derivatization
dc.typeArtículos de revistas


Este ítem pertenece a la siguiente institución