Artículos de revistas
Simultaneous Determination of Iron and Copper in Ethanol Fuel Using Nafion/Carbon Nanotubes Electrode
Fecha
2014-08-01Registro en:
Electroanalysis. Weinheim: Wiley-v C H Verlag Gmbh, v. 26, n. 8, p. 1794-1800, 2014.
1040-0397
10.1002/elan.201400136
WOS:000340557700018
0072173018005712
Autor
Universidade Estadual Paulista (Unesp)
Institución
Resumen
This work presents the analytical method development for iron and copper determination in ethanol fuel. This method was developed using stripping voltammetry with a glassy carbon electrode modified with Nafion/Carbon-nanotubes. With linear sweep stripping voltammetry was achieved a limit of detection of 7.1 x 10(-7) mol L-1 for Fe3+ and 5.1 x 10(-8) mol L-1 for Cu2+. The amperometric sensitivities were 2.0 x 10(6) mu A mol(-1) L for Fe3+ and 2.8 x 10(7) mu A mol(-1) L for Cu2+. The recovery study showed that the method has good accuracy and repeatability, with recovery of 108 and 103% for Fe3+ and Cu2+ respectively.