Artículos de revistas
A multicommuted flow-based system for hydrogen peroxide determination by chemiluminescence detection using a photodiode
Fecha
2007-01-01Registro en:
Analytical Letters, v. 40, n. 16, p. 3148-3157, 2007.
0003-2719
1532-236X
10.1080/00032710701603876
2-s2.0-38949175473
Autor
Universidade Federal da Bahia (UFBA)
Universidade Federal de São Carlos (UFSCar)
Universidade Estadual Paulista (Unesp)
Universidade de São Paulo (USP)
Indústria e Comércio FAC Ltda.
Institución
Resumen
A simple, rapid, and automated assay for hydrogen peroxide in pharmaceutical samples was developed by combining the multicommutation system with a chemiluminescence (CL) detector. The detection was performed using a spiral flow-cell reactor made from polyethylene tubing that was positioned in front of a photodiode. It allows the rapid mixing of CL reagent and analyte and simultaneous detection of the emitted light. The chemiluminescence was based on the reaction of luminol with hydrogen peroxide catalyzed by hexacyanoferrate(III). The feasibility of the flow system was ascertained by analyzing a set of pharmaceutical samples. A linear response within the range of 2.2-210 μmol l-1 H2O2 with a LD of 1.8 μmol l-1 H2O2 and coefficient of variations smaller than 0.8% for 1.0×10-5 mol l-1 and 6.8×10-5 mol l-1 hydrogen peroxide solutions (n=10) were obtained. Reagents consumption of 90 μg of luminol and 0.7 mg of hexacyanoferrate(III) per determination and sampling rate of 200 samples per hour were also achieved. Copyright © Taylor & Francis Group, LLC.