Actas de congresos
Screen-printed carbon electrode modified with poly-L-histidine applied to voltammetric determination of chromium (VI)
Fecha
2006-12-01Registro en:
ECS Transactions, v. 3, n. 10, p. 87-95, 2006.
1938-5862
1938-6737
10.1149/1.2357249
2-s2.0-33847015243
0000-0002-2296-1393
Autor
Universidade Estadual Paulista (Unesp)
Institución
Resumen
The present work reports the use of a screen-printed carbon electrode (SPCE) modified by poly-L-histidine film to determine chromium (VI). Stable films can be formed by direct addition of PH solution 1 % (w/v) on the electrode surface, followed by heating at 80°C during 5 min. Calibration curves can be constructed for Cr(VI) from 1.0 × 10-5 mol L-1 to 7.0 × 10-5 mol L-1 Cr (VI) in acetate buffer pH 4 using a preconcentration step of 60s at open circuit potential. A relative standard deviation of 3.2% was for five determination of 4.0 × 10 -5 mol L-1 Cr (VI). The method was successful applied to determination of Cr(VI) in wastewater samples from a leather dyeing industry. copyright The Electrochemical Society.